Validation of the U-STARR with the AB-York crescent of sound, a new instrument to evaluate speech intelligibility in noise and spatial hearing skills

Publication date

2015

Authors

Smulders, Yvette EISNI 000000038876746X
Rinia, Albert B.
Pourier, V. E.C.
Van Zon, Alice
van Zanten, Gijsbert A.
Stegeman, IngeORCID 0000-0001-5154-7178ISNI 0000000423219064
Scherf, Fanny W A C
Smit, Adriana LORCID 0000-0001-9126-9969
Topsakal, VedatISNI 0000000419547129
Tange, R. A.ISNI 0000000392752017

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

Abstract

The Advanced Bionics ® (AB)-York crescent of sound is a new test setup that comprises speech intelligibility in noise and localization tests that represent everyday listening situations. One of its tests is the Sentence Test with Adaptive Randomized Roving levels (STARR) with sentences and noise both presented from straight ahead. For the Dutch population, we adopted the AB-York setup and replaced the English sentences with a validated set of Dutch sentences. The Dutch version of the STARR is called the Utrecht-STARR (U-STARR). This study primarily assesses the validity and reliability of the U-STARR compared to the Plomp test, which is the current Dutch gold standard for speech-in-noise testing. The outcome of both tests is a speech reception threshold in noise (SRTn). Secondary outcomes are the SRTn measured with sounds from spatially separated sources (SISSS) as well as sound localization capability. We tested 29 normal-hearing adults and 18 postlingually deafened adult patients with unilateral cochlear implants (CI). This study shows that the U-STARR is adequate and reliable and seems better suited for severely hearing-impaired persons than the conventional Plomp test. Further, CI patients have poor spatial listening skills, as demonstrated with the AB-York test.

Keywords

Cochlear implant, Diagnostics, Hearing loss, Hearing test, Localization, Spatial hearing, Speech in noise, Otorhinolaryngology, Physiology, Speech and Hearing

Citation

Smulders, Y E, Rinia, A B, Pourier, V E C, Van Zon, A, Van Zanten, G A, Stegeman, I, Scherf, F W A C, Smit, A L, Topsakal, V, Tange, R A & Grolman, W 2015, 'Validation of the U-STARR with the AB-York crescent of sound, a new instrument to evaluate speech intelligibility in noise and spatial hearing skills', Audiology and Neurotology Extra, vol. 5, no. 1, pp. 1-10. https://doi.org/10.1159/000370300