Hypertensive Crisis During Norepinephrine Syringe Exchange: A Case Report

Publication date

2017-04

Authors

Snijder, Roland A
Knape, JTAISNI 0000000394537898
Egberts, Toine C.G.ORCID 0000-0003-1758-7779ISNI 0000000392745722
Timmerman, Annemoon M D EORCID 0000-0001-7421-0666ISNI 0000000391712718

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

Abstract

A 67-year critically ill patient suffered from a hypertensive crisis (200 mm Hg) because of a norepinephrine overdose. The overdose occurred when the clinician exchanged an almost-empty syringe and the syringe pump repeatedly reported an error. We hypothesized that an object between the plunger and the syringe driver may have caused the exertion of too much force on the syringe. Testing this hypothesis in vitro showed significant peak dosing errors (up to +572%) but moderate overdose (0.07 mL, +225%) if a clamp was used on the intravenous infusion line and a large overdose (0.8 mL, +2700%) if no clamp was used. Clamping and awareness are advised.

Keywords

General Medicine, Journal Article, Case Reports

Citation

Snijder, R A, Knape, J T A, Egberts, T C G & Timmerman, A M D E 2017, 'Hypertensive Crisis During Norepinephrine Syringe Exchange : A Case Report', Clinical case reports, vol. 8, no. 7, pp. 178-181. https://doi.org/10.1213/XAA.0000000000000458