Note: Rapid offset reduction of impedance bridges taking into account instrumental damping and phase shifting

Publication date

2013

Authors

van der Wel, C.M.
Kortschot, R.J.ISNI 0000000391296435
Bakelaar, I. A.ISNI 0000000419444314
Erné, B.ISNI 0000000397074702
Kuipers, BonnyORCID 0000-0002-0300-7190ISNI 0000000393666452

Editors

Advisors

Supervisors

Document Type

Article
Open Access logo

License

Abstract

The sensitivity of an imperfectly balanced impedance bridge is limited by the remaining offset voltage. Here, we present a procedure for offset reduction in impedance measurements using a lock-in amplifier, by applying a complex compensating voltage external to the bridge. This procedure takes into account instrumental damping and phase shifting, which generally occur at the high end of the operational frequency range. Measurements demonstrate that the output of the circuit rapidly converges to the instrumentally limited noise at any frequency.

Keywords

Citation

van der Wel, C M, Kortschot, R J, Bakelaar, I A, Erné, B H & Kuipers, B W M 2013, 'Note: Rapid offset reduction of impedance bridges taking into account instrumental damping and phase shifting', Review of Scientific Instruments, vol. 84, 036109, pp. 1-4. https://doi.org/10.1063/1.4795554