Diffusion weighted MRI with minimal distortion in head-and-neck radiotherapy using a turbo spin echo acquisition method

Publication date

2017

Authors

Schakel, TimORCID 0000-0002-1234-3479
Hoogduin, HansORCID 0000-0001-7707-0750ISNI 0000000389154198
Terhaard, Chris H.J.ORCID 0000-0001-6062-5457ISNI 0000000388691821
Philippens, Mariëlle E PORCID 0000-0003-0205-3459ISNI 0000000387302804

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

Abstract

PURPOSE: Diffusion weighted (DW) MRI, showing high contrast between tumor and background tissue, is a promising technique in radiotherapy for tumor delineation. However, its use for head-and-neck patients is hampered by poor geometric accuracy in conventional echo planar imaging (EPI) DW-MRI. An alternative turbo spin echo sequence, DW-SPLICE, is implemented and demonstrated in patients. METHODS: The DW-SPLICE sequence was implemented on a 3.0T system and evaluated in 10 patients. The patients were scanned in treatment position, using a customized head support and immobilization mask. Image distortions were quantified at the gross tumor volume (GTV) using field map analysis. The apparent diffusion coefficient (ADC) was evaluated using an ice water phantom. RESULTS: The DW images acquired by DW-SPLICE showed no image distortions. Field map analysis at the gross tumor volumes resulted in a median distortion of 0.2 mm for DW-SPLICE, whereas for the conventional method this was 7.2 mm. ADC values, measured using an ice water phantom were in accordance with literature values. CONCLUSIONS: The implementation of DW-SPLICE allows for diffusion weighted imaging of patients in treatment position with excellent geometrical accuracy. The images can be used to facilitate target volume delineation in RT treatment planning. This article is protected by copyright. All rights reserved.

Keywords

diffusion, head-and-neck cancer, image distortions, mri, radiotherapy, Journal Article

Citation

Schakel, T, Hoogduin, JM, Terhaard, C H J & Philippens, M E P 2017, 'Diffusion weighted MRI with minimal distortion in head-and-neck radiotherapy using a turbo spin echo acquisition method', Medical Physics, vol. 44, no. 8, pp. 4188-4193. https://doi.org/10.1002/mp.12363