Performance of a cylindrical diode array for use in a 1.5 T MR-linac

Publication date

2016-01-15

Authors

Houweling, A. C.ISNI 0000000387410098
De Vries, J. H W
Wolthaus, JWHISNI 0000000395891207
Woodings, Simon J
Kok, J. G MISNI 0000000391574140
van Asselen, BISNI 0000000391220676
Smit, K.
Bel, A.
Lagendijk, J J WISNI 0000000393637862
Raaymakers, Bas WORCID 0000-0002-8036-6808ISNI 0000000392005337

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

At the UMC Utrecht, a linear accelerator with integrated magnetic resonance imaging (MRI) has been developed, the MR-linac. Patient-specific quality assurance (QA) of treatment plans for MRI-based image guided radiotherapy requires QA equipment compatible with this 1.5 T magnetic field. The purpose of this study was to examine the performance characteristics of the ArcCHECK-MR in a transverse 1.5 T magnetic field. To this end, the short-term reproducibility, dose linearity, dose rate dependence, field size dependence, dose per pulse dependence and inter-diode dose response variation of the ArcCHECK-MR diode array were evaluated on a conventional linac and on the MR-linac. The ArcCHECK-MR diode array performed well for all tests on both linacs, no significant differences in performance characteristics were observed. Differences in the maximum dose deviations between both linacs were less than 1.5%. Therefore, we conclude that the ArcCHECK-MR can be used in a transverse 1.5 T magnetic field.

Keywords

ArcCHECK, ArcCHECK-MR, MR-linac, MRI-accelerator, patient specific QA, plan verification, radiotherapy, Taverne, Radiological and Ultrasound Technology, General Medicine, Radiology Nuclear Medicine and imaging, Journal Article

Citation

Houweling, A C, De Vries, J H W, Wolthaus, J, Woodings, S, Kok, J G M, Van Asselen, B, Smit, K, Bel, A, Lagendijk, J J W & Raaymakers, B W 2016, 'Performance of a cylindrical diode array for use in a 1.5 T MR-linac', Physics in Medicine and Biology, vol. 61, no. 3, pp. N80-N89. https://doi.org/10.1088/0031-9155/61/3/N80