MALDI mass spectrometry imaging in microscope mode with infrared lasers- bypassing the diffraction limits
Files
Publication date
2013
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
Abstract
This letter demonstrates the use of infrared matrix-assisted laser desorption/ionization coupled with microscope mode mass spectrometry imaging. It is aimed to explore the use of intrinsic water in tissue as a matrix for imaging at spatial resolutions below the diffraction limit of the employed IR optics. Stigmatic ion optics with a magnification factor of ∼70 were used to project the spatial distribution of produced ions onto a detector while separating ions with different mass-to-charge ratios using a time-offlight mass spectrometer. A pixelated detector was used to simultaneously record arrival time and impact position. A previously described dried-droplet sample system of 2,5- dihydroxybenzoic acid (DHB) and 5 peptides covered by a copper grid for defined surface structure was used to benchmark the light- and ion-optical setup for spatial resolution and mass spectrometric performance. A spatial resolving power of 9.8 μm, well below the optical limit of diffraction (14 μm for the given setup), was established. After, frozen cryosections from a biological model system were measured by exploiting the endogenous water content as a matrix. Principal component analysis enabled a clear distinction between distinct tissue regions identified by both light microscopy and MS imaging.
Keywords
Citation
Soltwisch, J, Göritz, G, Jungmann, JH, Kiss, A, Smith, D F, Ellis, S R & Heeren, R M A 2013, 'MALDI mass spectrometry imaging in microscope mode with infrared lasers- bypassing the diffraction limits', Analytical Chemistry, vol. 86, pp. 321-325. https://doi.org/10.1021/ac403421v