Dosimetry and linac acceptance for the 1.5 T MRI-Linac
Files
Publication date
2020-06-11
Authors
Woodings, Simon James
Editors
Advisors
Raaymakers, Bas W.
Asselen, B. van
Wolthaus, J.W.H.
Supervisors
DOI
Document Type
Dissertation
Metadata
Show full item recordCollections
License
Abstract
This thesis concerns the Unity MRI-linac (and its prototypes). The overall aim was to ensure that the MRI-linac was safe and effective for clinical use, and could be introduced into the clinic. To this end it was necessary to identify and solve issues related to reference dosimetry in a magnetic field, acceptance testing of a new system in a new physical configuration, and beam characterisation in a magnetic field.
Results and analysis are provided. A major contribution was the development of a reference dosimetry framework that allows dosimetry to be performed at an MRI-Linac by the local clinical physicist using standard equipment, and where all correction factors can be independently verified.
Additional contributions included evidence of the consistency of ion chamber magnetic field correction factors; evidence of the suitability of a microDiamond small-field detector; a linac acceptance procedure that can be performed by local clinical physicists, and; the first beam characterisation data suitable for modeling the radiation beam in the treatment planning system, such that the system could then be used to plan and deliver radiation treatments.
Keywords
MRI-Linac, MR-Linac, Unity, Dosimetry, Acceptance, Beam Characterization, kB, PTW microDiamond