A newly developed tool for intra-tracheal temperature and humidity assessment in laryngectomized individuals: the Airway Climate Explorer (ACE)
Publication date
2007
Authors
Zuur, J.K.
Muller, S.H.
Jongh, F.H.C.
Horst, M.J. van der
Shehata, M.
Leeuwen, J. van
Sinaasappel, M.
Hilgers, F.J.M.
Editors
Advisors
Supervisors
DOI
Document Type
Article
Metadata
Show full item recordCollections
License
Abstract
The aim of this study is to develop a postlaryngectomy airway climate explorer (ACE) for assessment of intratracheal temperature and humidity and of influence of heat and moisture exchangers (HMEs). Engineering goals were within-device condensation prevention and fast response time characteristics. The ACE consists of a small diameter, heated air-sampling catheter connected to a heated sensor house, containing a humidity sensor. Air is sucked through the catheter by a controlled-flow pump. Validation was performed in a climate chamber using a calibrated reference sensor and in a two-flow system. Additionally, the analyser was tested in vivo. Over the clinically relevant range of humidity values (5–42 mg H2O/l air) the sensor output highly correlates with the reference sensor readings (R 2 > 0.99). The 1–1/e response times are all <0.5 s. A first in vivo pilot measurement was successful. The newly developed, verified, fast-responding ACE is suitable for postlaryngectomy airway climate assessment.
Keywords
Gas sampling probe, Heat and moisture exchanger, Airway climate, Temperature, Humidity