Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study

Publication date

2017

Authors

van Hamersvelt, Robbert W
Schilham, Arnold M RISNI 0000000391935778
Engelke, Klaus
den Harder, Annemarie M.
de Keizer, BartORCID 0000-0002-6270-9483ISNI 0000000393842428
Verhaar, Harald J
Leiner, TimORCID 0000-0003-1885-5499ISNI 0000000390698205
de Jong, PimORCID 0000-0003-4840-6854ISNI 0000000395539334
Willemink, Martin J.

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

Abstract

Objectives: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. Methods: Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm 3 calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90–200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA). Results: At both 120 and 140 kVp, excellent correlations (R = 0.97, P < 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm 3 (-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm 3 (-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R 2 ≥ 0.970, P < 0.001) with DXA were found. Conclusions: SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings. Key points: • Dual-layer spectral detector CT allows for accurate bone mineral density quantification. • BMD measurements on SDCT are strongly linearly correlated to DXA. • SDCT, acquired for several indications, may allow for evaluation of osteoporosis. • This potentially opens up the possibility for opportunistic osteoporosis screening.

Keywords

Dual energy X-Ray absorptiometry, Bone mineral density, Dual-energy CT, Dual-layer spectral detector CT, Material decomposition, Journal Article

Citation

van Hamersvelt, R W, Schilham, A M R, Engelke, K, den Harder, A M, de Keizer, B, Verhaar, H J, Leiner, T, de Jong, P A & Willemink, M J 2017, 'Accuracy of bone mineral density quantification using dual-layer spectral detector CT : a phantom study', European Radiology, vol. 27, no. 10, pp. 4351-4359. https://doi.org/10.1007/s00330-017-4801-4