Irradiance levels of phototherapy devices: A national study in Dutch neonatal intensive care units
Publication date
2017-07-01
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
taverne
Abstract
Objective: The objective of this study is to determine whether irradiance levels of phototherapy (PT) devices in Dutch neonatal intensive care units (NICUs) increased between 2008 and 2013. Study Design: Irradiance of all types of PT devices, used in combination with incubators, was measured with a Dale 40 Radiometer (Fluke Biomedical, Everett, WA, USA) in all 10 Dutch NICUs. Results: Irradiance increased in seven NICUs. Median (range) irradiance increased from 9.7 (4.3-32.6) to 16.4 (6.8-41) μW cm-2 nm-1 for 24 overhead devices (P=0.004) and from 6.8 (0.8-15.6) to 22.3 (1.1-36.3) μW cm-2 nm-1 for 12 underneath devices (P=0.014). Five light-emitting diode (LED)-based devices were used in 2013 and one in 2008. The mean distance between overhead PT device and infant decreased by ∼9 cm (P<0.001). Significantly more devices delivered minimal (10 μW cm-2 nm-1) recommended irradiance levels (80 vs ∼45%; P=0.002). Conclusion: Irradiance of PT devices still varies, but has markedly improved since 2008 due to shorter distances between PT device and infant, and introduction of better performing LED-based devices.
Keywords
Taverne, Pediatrics, Perinatology, and Child Health, Obstetrics and Gynaecology, Journal Article
Citation
Hulzebos, C V, Van't Klooster, S J, Krummel-Lorenz, B, Vreman, H J, Dijk, P H, Benders, M J N L, Buijs, J, Dudink, J, Van Imhoff, D E, Laarman, C R, Liem, K D, van Lingen, R A, Lopriore, E, Mulder, A L M, Toledo-Eppinga, L & BARTrial Study Group 2017, 'Irradiance levels of phototherapy devices : A national study in Dutch neonatal intensive care units', Journal of perinatology, vol. 37, no. 7, pp. 839-842. https://doi.org/10.1038/jp.2017.13