Ex vivo sealing performance of a sutureless dual-component connector for coronary bypass

Publication date

2025-11-01

Authors

Gianoli, MonicaORCID 0000-0002-8835-2339
Wassink, Harmen M
van Schelven, Leonard JORCID 0000-0001-7903-0306ISNI 0000000390058494
Grundeman, PaulISNI 0000000392925950
Suyker, Willem J LISNI 0000000394629063

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

Sutureless coronary connectors may enable minimally invasive bypass surgery, but consistent sealing in human vessels remains a key challenge. This ex vivo study evaluates the hydrostatic performance of a titanium connector that employs a novel sealing principle: tissue-to-metal coaptation enabling stepwise distal anastomosis construction. Thirteen anastomoses were created in porcine and human cadaveric vessels using a standardized three-step deployment protocol. Target sites included internal mammary arteries, saphenous veins, and coronary arteries, selected for suitability for hand-suturing. Sealing was assessed by hydrostatic pressure testing up to 300 mmHg. Key anatomical parameters and procedural outcomes were recorded. Twelve of 13 anastomoses were completed without leakage up to 300 mmHg; one case exhibited minor oozing at 260 mmHg. The connector accommodated vessels down to 1.3 mm in diameter and up to 0.8 mm wall thickness. One misdeployment due to an oversized incision was corrected intraoperatively without compromising the hydrostatic result. Visual inspection confirmed wide, unobstructed anastomotic orifices. This ex vivo evaluation demonstrates that the connector achieved consistent hydrostatic sealing through bare-metal vessel coaptation, even in small-diameter human coronary tissue. These findings support further in vivo validation towards minimally invasive coronary bypass applications.

Keywords

anastomotic connector device, coronary revascularization, vascular connector technology, Surgery, Pulmonary and Respiratory Medicine, Cardiology and Cardiovascular Medicine

Citation

Gianoli, M, Wassink, H M, van Schelven, L J, Gründeman, P F & Suyker, W J L 2025, 'Ex vivo sealing performance of a sutureless dual-component connector for coronary bypass', Interdisciplinary cardiovascular and thoracic surgery, vol. 40, no. 11, ivaf270. https://doi.org/10.1093/icvts/ivaf270