Development and validation of an HPLC-MS/MS method to quantify the KRAS inhibitor adagrasib in mouse plasma and tissue-related matrices
Publication date
2023-11
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by_nc_nd
Abstract
We developed and validated an assay utilizing a liquid chromatography-tandem mass spectrometry technique to quantify the KRAS inhibitor adagrasib in mouse plasma and seven tissue-related matrices. The straightforward protein precipitation technique was selected to extract adagrasib and the internal standard salinomycin from the matrices. Gradient elution of acetonitrile and water modified with 0.5% (v/v) ammonium hydroxide and 0.02% (v/v) acetic acid on a C 18 column at a flow rate of 0.6 ml/min was applied to separate the analytes. Both adagrasib and salinomycin were detected with a triple quadrupole mass spectrometer with positive electrospray ionization in a selected reaction monitoring mode. A linear calibration range of 2-2,000 ng/ml of adagrasib was demonstrated during the validation. In addition, the reported precision values (intra- and inter-day) were between 3.5 and 14.9%, while the accuracy values were 85.5-111.0% for all tested levels in all investigated matrices. Adagrasib in mouse plasma was reported to have good stability at room temperature, while adagrasib in tissue-related matrices was stable on ice for up to 4 h (matrix dependent). Finally, this method was successfully applied to determine the pharmacokinetic profile and tissue distribution of adagrasib in wild-type mice.
Keywords
KRAS inhibitor, LC–MS/MS, adagrasib, bioanalysis, mouse matrices, Drug Discovery, Analytical Chemistry, Molecular Biology, Biochemistry, Clinical Biochemistry, Pharmacology
Citation
Retmana, I A, Loos, N H C, Schinkel, A H, Beijnen, J H & Sparidans, R W 2023, 'Development and validation of an HPLC-MS/MS method to quantify the KRAS inhibitor adagrasib in mouse plasma and tissue-related matrices', Biomedical Chromatography, vol. 37, no. 11, e5720. https://doi.org/10.1002/bmc.5720