Comparison of septal strain patterns in dyssynchronous heart failure between speckle tracking echocardiography vendor systems

Publication date

2015-01-27

Authors

van Everdingen, Wouter MISNI 0000000464452962
Paiman, Marie-Louise
van Deursen, Caroline J M
Cramer, Maarten JISNI 0000000390984527
Vernooy, Kevin
Delhaas, Tammo
Prinzen, Frits W

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

AIM: To analyze inter-vendor differences of speckle tracking echocardiography (STE) in imaging cardiac deformation in patients with dyssynchronous heart failure. METHODS AND RESULTS: Eleven patients (all with LBBB, median age 60.7 years, 9 males) with implanted cardiac resynchronization therapy devices were prospectively included. Ultrasound systems of two vendors (i.e. General Electric and Philips) were used to record images in the apical four chamber view. Regional longitudinal strain patterns were analyzed with vendor specific software in the basal, mid and apical septal segments. Systolic strain (SS), time to peak strain (TTP) and septal rebound stretch (SRS) were determined during four pacing settings, resulting in 44 unique strain patterns per segment (total 132 patterns). Cross correlation was used to analyze the comparability of the shape of 132 normalized strain patterns. Correlation of strain patterns of the two systems was high (R(2) median: 0.68, interquartile range: 0.53-0.82). Accordingly, strain patterns of intrinsic rhythm were recognized equally using both systems, when divided into three types. GE based SS (18.9 ± 4.7%) was significantly higher than SS determined by the Philips system (13.4 ± 4.3%). TTP was slightly but non-significantly lower in GE (384 ± 77 ms) compared to Philips (404 ± 83 ms) derived strain signals. Correlation of SRS between the systems was poor, due to minor differences in the strain signal and timing of aortic valve closure. CONCLUSIONS: The two systems provide similar shape of strain patterns. However, important differences are found in the amplitude, timing of systole and SRS. Until STE is standardized, clinical decision making should be restricted to pattern analysis.

Keywords

Taverne

Citation

van Everdingen, W M, Paiman, M-L, van Deursen, C J M, Cramer, MJ, Vernooy, K, Delhaas, T & Prinzen, F W 2015, 'Comparison of septal strain patterns in dyssynchronous heart failure between speckle tracking echocardiography vendor systems', Journal of Electrocardiology, vol. 48, no. 4, pp. 609-16. https://doi.org/10.1016/j.jelectrocard.2014.12.021