A high-precision network analyzer system for the measurement of the electrode impedance of both stationary and non-stationary electrode, with special attention to the dropping mercury electrode
Publication date
1980-05-26
Authors
Bongenaar, C.P.M.
Sluyters-Rehbach, M.
Sluyters, J.H.
Editors
Advisors
Supervisors
DOI
Document Type
Article
Metadata
Show full item recordCollections
License
Abstract
A largely automatic instrument for the measurement of the admittance of a galvanic cell is described. During a measurement the frequency of the alternating voltage and the dc voltage are scanned stepwise in a programmed sequence. A high precision has been obtained by applying a programmed correction procedure and an automatic correction for short-term drift.
The performance of the instrument is demonstrated on a dummy cell, a cell with supporting electrolyte only and on the determination of the kinetic parameters of the fast Cd(II) ion reduction on the DME in 1 M KCl.