Reducing distortions in echo-planar breast imaging at ultrahigh field with high-resolution off-resonance maps

Publication date

2019-07

Authors

van Rijssel, Michael J.
Zijlstra, Frank
Seevinck, P.R.ISNI 0000000390489892
Luijten, Peter R.ORCID 0000-0002-8040-8449ISNI 0000000397136870
Gilhuijs, KennethORCID 0000-0003-2087-8649ISNI 0000000393336330
Klomp, Dennis W.J.ORCID 0000-0002-5884-5386ISNI 0000000396514008
Pluim, Josien P WORCID 0000-0001-7327-9178ISNI 000000014097262X

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by_nc

Abstract

Purpose: DWI is a promising modality in breast MRI, but its clinical acceptance is slow. Analysis of DWI is hampered by geometric distortion artifacts, which are caused by off-resonant spins in combination with the low phase-encoding bandwidth of the EPI sequence used. Existing correction methods assume smooth off-resonance fields, which we show to be invalid in the human breast, where high discontinuities arise at tissue interfaces. Methods: We developed a distortion correction method that incorporates high-resolution off-resonance maps to better solve for severe distortions at tissue interfaces. The method was evaluated quantitatively both ex vivo in a porcine tissue phantom and in vivo in 5 healthy volunteers. The added value of high-resolution off-resonance maps was tested using a Wilcoxon signed rank test comparing the quantitative results obtained with a low-resolution off-resonance map with those obtained with a high-resolution map. Results: Distortion correction using low-resolution off-resonance maps corrected most of the distortions, as expected. Still, all quantitative comparison metrics showed increased conformity between the corrected EPI images and a high-bandwidth reference scan for both the ex vivo and in vivo experiments. All metrics showed a significant improvement when a high-resolution off-resonance map was used (P < 0.05), in particular at tissue boundaries. Conclusion: The use of off-resonance maps of a resolution higher than EPI scans significantly improves upon existing distortion correction techniques, specifically by superior correction at glandular tissue boundaries.

Keywords

7T, B mapping, breast, diffusion-weighted imaging, distortion correction, echo-planar imaging, off-resonance, Radiology Nuclear Medicine and imaging, Journal Article

Citation

van Rijssel, M J, Zijlstra, F, Seevinck, P R, Luijten, P R, Gilhuijs, K G A, Klomp, D W J & Pluim, J P W 2019, 'Reducing distortions in echo-planar breast imaging at ultrahigh field with high-resolution off-resonance maps', Magnetic Resonance in Medicine, vol. 82, no. 1, pp. 425-435. https://doi.org/10.1002/mrm.27701