On the suitability of Elekta's Agility 160 MLC for tracked radiation delivery: closed-loop machine performance

Publication date

2015-03-07

Authors

Glitzner, M
Crijns, S P MISNI 0000000396863953
de Senneville, B. Denis
Lagendijk, J J WISNI 0000000393637862
Raaymakers, BasORCID 0000-0002-8036-6808ISNI 0000000392005337

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

For motion adaptive radiotherapy, dynamic multileaf collimator tracking can be employed to reduce treatment margins by steering the beam according to the organ motion. The Elekta Agility 160 MLC has hitherto not been evaluated for its tracking suitability. Both dosimetric performance and latency are key figures and need to be assessed generically, independent of the used motion sensor. In this paper, we propose the use of harmonic functions directly fed to the MLC to determine its latency during continuous motion. Furthermore, a control variable is extracted from a camera system and fed to the MLC. Using this setup, film dosimetry and subsequent. statistics are performed, evaluating the response when tracking (MRI)-based physiologic motion in a closed-loop. The delay attributed to the MLC itself was shown to be a minor contributor to the overall feedback chain as compared to the impact of imaging components such as MRI sequences. Delay showed a linear phase behaviour of the MLC employed in continuously dynamic applications, which enables a general MLC-characterization. Using the exemplary feedback chain, dosimetry showed a vast increase in pass rate employing. statistics. In this early stage, the tracking performance of the Agility using the test bench yielded promising results, making the technique eligible for translation to tracking using clinical imaging modalities.

Keywords

MLC, tracking, system response, delay analysis, band-limited, linear-phase, physiologic motion, PROOF-OF-CONCEPT, TARGET TRACKING, MOTION, RADIOTHERAPY, Taverne

Citation

Glitzner, M, Crijns, S P M, de Senneville, B D, Lagendijk, J J W & Raaymakers, B W 2015, 'On the suitability of Elekta's Agility 160 MLC for tracked radiation delivery : closed-loop machine performance', Physics in Medicine and Biology, vol. 60, no. 5, pp. 2005-2017. https://doi.org/10.1088/0031-9155/60/5/2005