Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study
Publication date
2021-11-24
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by
Abstract
OBJECTIVE: To quantify metal artifact reduction using 130 keV virtual monochromatic imaging (VMI) with and without orthopedic metal artifact reduction (O-MAR) in total hip arthroplasty. METHODS: Conventional polychromatic images and 130 keV VMI of a phantom with pellets representing bone with unilateral or bilateral prostheses were reconstructed with and without O-MAR on a dual-layer CT. Pellets were categorized as unaffected, mildly affected and severely affected. RESULTS: When 130 keV VMI with O-MAR was compared to conventional imaging with O-MAR, a relative metal artifact reduction in CT values, contrast-to-noise (CNR), signal-to-noise (SNR) and noise in mildly affected pellets (67%, 74%, 48%, 68%, respectively; p < 0.05) was observed but no significant relative metal artifact reduction in severely affected pellets. Comparison between 130 keV VMI without O-MAR and conventional imaging with O-MAR showed relative metal artifact reduction in CT values, CNR, SNR and noise in mildly affected pellets (92%, 72%, 38%, 51%, respectively; p < 0.05) but negative relative metal artifact reduction in CT values and noise in severely affected pellets (- 331% and -223%, respectively; p < 0.05), indicating aggravation of metal artifacts. CONCLUSION: Overall, VMI of 130 keV with O-MAR provided the strongest metal artifact reduction.
Keywords
Arthroplasty, CT, Dual energy, Hip, Replacement, Radiology Nuclear Medicine and imaging
Citation
Selles, M, Stuivenberg, V H, Wellenberg, R H H, van de Riet, L, Nijholt, I M, van Osch, J A C, van Hamersvelt, R W, Leiner, T & Boomsma, M F 2021, 'Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study', Insights into Imaging, vol. 12, no. 1, 171. https://doi.org/10.1186/s13244-021-01111-5