Safety and effectiveness of the first balloon-in-basket pulsed field ablation system for the treatment of atrial fibrillation: VOLT CE Mark Study 6-month results
Files
Publication date
2025-04-01
Authors
Tilz, Roland R.
Chierchia, Gian Battista
Gunawardene, Melanie
Sanders, Prashanthan
Haqqani, Haris
Kalman, Jonathan
Healy, Stewart
Pürerfellner, Helmut
Neuzil, Petr
Asensi, Joaquín Osca
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by
Abstract
Aims Pulsed field ablation (PFA) is a growing ablation modality for pulmonary vein isolation (PVI) in atrial fibrillation (AF) patients. This study assesses the 6-month safety and effectiveness of a novel balloon-in-basket, mapping-integrated PFA system, with a purpose-built form factor for PVI. Methods and results The VOLT CE Mark Study is a prospective, multi-center, pre-market study. A total of 150 patients with drug-refractory paroxysmal (PAF) or persistent AF (PersAF) were enrolled between 8 November 2023 and 14 March 2024, of which 146 patients (age 64.1 ± 10.0 years, 63.0% male, 70.5% PAF) underwent PVI with the balloon-in-basket PFA catheter and system featuring integrated electroanatomic mapping with contact-sensing. Study endpoints were the rate of primary serious adverse events within 7 days as well as acute procedural effectiveness and 6-month freedom from recurrence. Acute effectiveness was achieved in 99.1% (573/578) of treated PVs (98.6% of patients, 144/146) with 17.6 ± 5.7 PFA applications/patient. Procedure, fluoroscopy, LA dwell, and transpired ablation times were 100.4 ± 33.0, 17.3 ± 12.1, 39.4 ± 20.6, and 31.4 ± 16.8 min, respectively. There were 4 (2.7%; 4/146) primary serious adverse events. The rate of freedom from documented atrial arrhythmias was 88.2% in PAF patients and 76.7% in PersAF patients (freedom from symptomatic recurrence was documented in 90.2% of PAF patients and 74.4% of PersAF patients) through 6-months post-index procedure. Conclusion The VOLT CE Mark Study primary results demonstrate the safety and effectiveness of the novel balloon-in-basket PFA system to perform PVI in PAF and PersAF.
Keywords
Atrial fibrillation, Balloon, Balloon-in-basket, Basket, Catheter ablation, Pulmonary vein isolation, Pulsed field ablation, Single-shot, Cardiology and Cardiovascular Medicine, Physiology (medical)
Citation
Tilz, R R, Chierchia, G B, Gunawardene, M, Sanders, P, Haqqani, H, Kalman, J, Healy, S, Pürerfellner, H, Neuzil, P, Asensi, J O, Loh, P, Reddy, V Y, Knecht, S, Jesser, E, Dirckx, N, Miller, A, Walker, D & Lakkireddy, D 2025, 'Safety and effectiveness of the first balloon-in-basket pulsed field ablation system for the treatment of atrial fibrillation : VOLT CE Mark Study 6-month results', Europace, vol. 27, no. 4, euaf072. https://doi.org/10.1093/europace/euaf072