Robust filtering of thin-slice reconstructions improves lacunar stroke detection in CT perfusion imaging

Publication date

2026-01

Authors

Vromans, Joris
Bennink, EdwinORCID 0000-0002-3689-8532ISNI 0000000419549773
Dankbaar, Jan WillemISNI 0000000392895296
Velthuis, BirgittaORCID 0000-0002-2542-9474ISNI 0000000395231874
de Jong, Hugo W A MORCID 0000-0002-3000-8316

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

BACKGROUND: The detection of lacunar strokes in CT perfusion (CTP) imaging may be improved by reducing the reconstructed CTP slice thickness. However, this increases noise, necessitating the use of robust noise filters. PURPOSE: To Investigate the impact of slice thickness and three state-of-the-art noise filters on the detection of lacunar strokes of varying sizes. METHODS: Artificial spherical lacunar infarcts (diameter: 5, 7, and 10 mm) were added to 40 thin slice CTP acquisitions. The performance of time-intensity profile guided filtering, temporal-average guided filtering, and a U-Net based filter were compared based on infarct detectability in cerebral blood volume maps of thick- (4.9 mm) and thin-slice (0.7 mm) reconstructions. One observer evaluated the maps for each combination of filter and slice thickness. A second observer repeated the evaluation for the two best performing combinations. F1-score, contrast-to-noise ratio, and observer confidence were calculated for every combination of filter, slice thickness, and observer. RESULTS: The highest F1-scores were achieved with the temporal-average guided filter (thin slices: 0.70, thick slices: 0.49). The other filters did not exceed an F1-score of 0.36 on either thin- or thick slices. Thin-slice temporal-average guided filtering also achieved the highest contrast-to-noise ratio and observer confidence, which resulted in a sensitivity of 10% for 5 mm, 73% for 7 mm, and 75% for 10 mm diameter lacunar infarcts. CONCLUSIONS: This study demonstrates that robust noise filtering, particularly temporal-average guided filtering, enables improved detection of lacunar strokes (≥7 mm) in CTP imaging using thin-slice reconstructions.

Keywords

Humans, Image Processing, Computer-Assisted/methods, Perfusion Imaging/methods, Signal-To-Noise Ratio, Stroke, Lacunar/diagnostic imaging, Tomography, X-Ray Computed/methods, Journal Article

Citation

Vromans, J, Bennink, E, Dankbaar, J W, Velthuis, B K & de Jong, H W A M 2026, 'Robust filtering of thin-slice reconstructions improves lacunar stroke detection in CT perfusion imaging', Medical physics, vol. 53, no. 1, e70271. https://doi.org/10.1002/mp.70271