Removal of urea by electro-oxidation in a miniature dialysis device: a study in awake goats

Publication date

2018-11

Authors

Wester, Maarten
van Gelder, M.
Joles, Jaap A.ORCID 0000-0003-2565-242XISNI 0000000396018725
Simonis, Frank F.J.
Hazenbrink, Diënty H M
van Berkel, Theo W M
Vaessen, Koen R D
Boer, Walther H
Verhaar, Marianne C.ORCID 0000-0002-3276-6428ISNI 0000000390259392
Gerritsen, KarinORCID 0000-0002-1917-6255ISNI 0000000388831038

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

The key to success in developing a wearable dialysis device is a technique to safely and efficiently regenerate and reuse a small volume of dialysate in a closed-loop system. In a hemodialysis model in goats, we explored whether urea removal by electro-oxidation (EO) could be effectively and safely applied in vivo. A miniature dialysis device was built, containing 1 or 2 "EO units," each with 10 graphite electrodes, with a cumulative electrode surface of 585 cm2 per unit. The units also contained poly(styrene-divinylbenzene) sulfonate beads, FeOOH beads, and activated carbon for respective potassium, phosphate, and chlorine removal. Urea, potassium, and phosphate were infused to create "uremic" conditions. Urea removal was dependent on total electrode surface area [removal of 8 mmol/h (SD 1) and 16 mmol/h (SD 2) and clearance of 12 ml/min (SD 1) and 20 ml/min (SD 3) with 1 and 2 EO units, respectively] and plasma urea concentration but not on flow rate. Extrapolating urea removal with 2 EO units to 24 h would suffice to remove daily urea production, but for intermittent dialysis, additional units would be required. EO had practically no effects on potassium and phosphate removal or electrolyte balance. However, slight ammonium releasewas observed, and some chlorine release at higher dialysate flow rates. Minor effects on acid-base balance were observed, possibly partly due to infusion of chloride. Mild hemolysis occurred, which seemed related to urea infusion. In conclusion, clinically relevant urea removal was achieved in vivo by electro-oxidation. Efficacy and safety testing in a large-animal model with uremia is now indicated.

Keywords

Acid-Base Equilibrium, Acid-Base Imbalance/etiology, Animals, Creatinine/blood, Dialysis Solutions/metabolism, Disease Models, Animal, Equipment Design, Goats, Hemolysis, Journal Article, Miniaturization, Models, Biological, Oxidation-Reduction, Phosphates/blood, Potassium/blood, Renal Dialysis/adverse effects, Time Factors, Urea/blood, Uremia/blood, Wakefulness, Wearable Electronic Devices, electro-oxidation, hemodialysis, in vivo, medical device, urea, wearable, Urology, Physiology, Research Support, Non-U.S. Gov't, Journal Article

Citation

Wester, M, van Gelder, M K, Joles, J A, Simonis, F, Hazenbrink, D H M, van Berkel, T W M, Vaessen, K R D, Boer, W H, Verhaar, M C & Gerritsen, K G F 2018, 'Removal of urea by electro-oxidation in a miniature dialysis device : a study in awake goats', American Journal of Physiology-Renal Physiology, vol. 315, no. 5, pp. F1385-F1397. https://doi.org/10.1152/ajprenal.00094.2018