Treatment planning evaluation and experimental validation of the magnetic resonance-based intrafraction drift correction

Publication date

2024-04

Authors

van den Dobbelsteen, Madelon
Hackett, Sara L
van Asselen, BISNI 0000000391220676
Oolbekkink, Stijn
Raaymakers, Bas WORCID 0000-0002-8036-6808ISNI 0000000392005337
de Boer, Johannes C J

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

Background and purpose: MRI-guided online adaptive treatments can account for interfractional variations, however intrafraction motion reduces treatment accuracy. Intrafraction plan adaptation methods, such as the Intrafraction Drift Correction (IDC) or sub-fractionation, are needed. IDC uses real-time automatic monitoring of the tumor position to initiate plan adaptations by repositioning segments. IDC is a fast adaptation method that occurs only when necessary and this method could enable margin reduction. This research provides a treatment planning evaluation and experimental validation of the IDC. Materials and methods: An in silico treatment planning evaluation was performed for 13 prostate patients mid-treatment without and with intrafraction plan adaptation (IDC and sub-fractionation). The adaptation methods were evaluated using dose volume histogram (DVH) metrics. To experimentally verify IDC a treatment was mimicked whereby a motion phantom containing an EBT3 film moved mid-treatment, followed by repositioning of segments. In addition, the delivered treatment was irradiated on a diode array phantom for plan quality assurance purposes. Results: The planning study showed benefits for using intrafraction adaptation methods relative to no adaptation, where the IDC and sub-fractionation showed consistently improved target coverage with median target coverages of 100.0%. The experimental results verified the IDC with high minimum gamma passing rates of 99.1% and small mean dose deviations of maximum 0.3%. Conclusion: The straightforward and fast IDC technique showed DVH metrics consistent with the sub-fractionation method using segment weight re-optimization for prostate patients. The dosimetric and geometric accuracy was shown for a full IDC workflow using film and diode array dosimetry.

Keywords

Intrafraction drift correction, Intrafraction motion, Motion management, MR-guided radiotherapy, Radiation, Oncology, Radiology Nuclear Medicine and imaging

Citation

van den Dobbelsteen, M, Hackett, S L, van Asselen, B, Oolbekkink, S, Raaymakers, B W & de Boer, J C J 2024, 'Treatment planning evaluation and experimental validation of the magnetic resonance-based intrafraction drift correction', Physics and Imaging in Radiation Oncology, vol. 30, 100580. https://doi.org/10.1016/j.phro.2024.100580