31 P MR Spectroscopy in the Pancreas: Repeatability, Comparison With Liver, and Pilot Pancreatic Cancer Data.

Publication date

2024-12

Authors

Seelen, Leonard W F
van den Wildenberg, Lieke
Gursan, Ayhan
Froeling, MartijnORCID 0000-0003-3841-0497
Gosselink, Mark W J M
van der Kemp, Wybe J. M.ISNI 0000000397017687
Mohammad, Nadia HajORCID 0000-0002-4688-2921
Molenaar, I QuintusORCID 0000-0002-1585-7184ISNI 0000000107493758
van Santvoort, Hjalmar CISNI 0000000389663785
Klomp, Dennis W.J.ORCID 0000-0002-5884-5386ISNI 0000000396514008

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by_nc_nd

Abstract

Background: Non-invasive evaluation of phosphomonoesters (PMEs) and phosphodiesters (PDEs) by 31-phosphorus MR spectroscopy (31P MRS) may have potential for early therapy (non-)response assessment in cancer. However, 31P MRS has not yet been applied to investigate the human pancreas in vivo. Purpose: To assess the technical feasibility and repeatability of 31P MR spectroscopic imaging (MRSI) of the pancreas, compare 31P metabolite levels between pancreas and liver, and determine the feasibility of 31P MRSI in pancreatic cancer. Study Type: Prospective cohort study. Population: 10 healthy subjects (age 34 ± 12 years, four females) and one patient (73-year-old female) with pancreatic ductal adenocarcinoma. Field Strength/Sequence: 7-T, 31P FID-MRSI, 1H gradient-echo MRI. Assessment: 31P FID-MRSI of the abdomen (including the pancreas and liver) was performed with a nominal voxel size of 20 mm (isotropic). For repeatability measurements, healthy subjects were scanned twice on the same day. The patient was only scanned once. Test–retest 31P MRSI data of pancreas and liver voxels (segmented on 1H MRI) of healthy subjects were quantified by fitting in the time domain and signal amplitudes were normalized to γ-adenosine triphosphate. In addition, the PME/PDE ratio was calculated. Metabolite levels were averaged over all voxels within the pancreas, right liver lobe and left liver lobe, respectively. Statistical Tests: Repeatability of test–retest data from healthy pancreas was assessed by paired t-tests, Bland–Altman analyses, and calculation of the intrasubject coefficients of variation (CoVs). Significant differences between healthy pancreas and right and left liver lobes were assessed with a two-way analysis of variance (ANOVA) for repeated measures. A P-value <0.05 was considered statistically significant. Results: The intrasubject CoVs for PME, PDE, and PME/PDE in healthy pancreas were below 20%. Furthermore, PME and PME/PDE were significantly higher in pancreas compared to liver. In the patient with pancreatic cancer, qualitatively, elevated relative PME signals were observed in comparison with healthy pancreas. Data Conclusion: In vivo 31P MRSI of the human healthy pancreas and in pancreatic cancer may be feasible at 7 T. Evidence Level: 3. Technical Efficacy: Stage 2.

Keywords

P MR spectroscopy, liver, metabolites, pancreas, pancreatic cancer, repeatability, Radiology Nuclear Medicine and imaging, Journal Article

Citation

Seelen, L W F, van den Wildenberg, L, Gursan, A, Froeling, M, Gosselink, M W J M, van der Kemp, W J M, Haj Mohammad, N, Molenaar, I Q, van Santvoort, H C, Klomp, D W J & Prompers, J J 2024, '31 P MR Spectroscopy in the Pancreas : Repeatability, Comparison With Liver, and Pilot Pancreatic Cancer Data.', Journal of Magnetic Resonance Imaging, vol. 60, no. 6, pp. 2657-2666. https://doi.org/10.1002/jmri.29326