Verification Study of Residual Activity Measurements After Yttrium-90 Radioembolization with Glass Microspheres
Publication date
2020-09
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
Abstract
OBJECTIVE: After yttrium-90 (90Y) radioembolization, residual activity and its consequences for dosimetric calculations are often not reported. The manufacturer for glass microspheres prescribes standard residual activity measurements by a survey meter, but the validity lacks evidence. This study aims to verify the accuracy of the survey meter approach for measuring residual activity of glass microspheres after treatment with glass microspheres. METHODS: To validate the accuracy of the survey meter approach, the measured residual activity of glass microspheres by survey meter was compared with measurements by PET. A sample of these waste containers was also measured by dose calibrator to confirm the accuracy of the PET. RESULTS: Twenty-four waste containers from glass microsphere treatments were prospectively scanned with 90Y-PET/CT. Bland-Altman plots showed substantial disagreement in residual activity measured by survey meter versus the residual activity measured by PET and dose calibrator, whereas the correlation between PET and dose calibrator was excellent (ρ = 0.99). CONCLUSION: This study found a significant disagreement between the residual activities measured by the survey meter, compared to measurements by PET and dose calibrator. If relatively high amounts of residual activity are encountered using the exposure rate measurement with a survey meter, additional quantification should be considered using either PET/CT or a dose calibrator measurement.
Keywords
s Yttrium-90 microsphere, Radioembolization, SIRT, TheraSphere, Residual activity, Survey meter, Yttrium-90 microsphere, Cardiology and Cardiovascular Medicine, Radiology Nuclear Medicine and imaging, Journal Article
Citation
Ebbers, S C, Kunnen, B, van Nierop, B J, Bemelmans, J L M, Krijger, G C, Lam, M G E H & Braat, A J A T 2020, 'Verification Study of Residual Activity Measurements After Yttrium-90 Radioembolization with Glass Microspheres', Cardiovascular and Interventional Radiology, vol. 43, no. 9, pp. 1378-1383. https://doi.org/10.1007/s00270-020-02504-7