Svelte Integrated Delivery System Performance Examined Through Diagnostic Catheter Delivery: The SPEED Registry

Publication date

2015-01-01

Authors

Khattab, Ahmed A.
Nijhoff, Freek
Schofer, Joachim
Berland, Jacques
Meier, Bernhard
Nietlispach, Fabian
Agostoni, PiefrancescoISNI 0000000388177910
Brucks, Steffen
Stella, Pieter R.ISNI 0000000393574363

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

Aims: The multi-center SPEED registry evaluated the procedural success and in-hospital clinical outcomes of direct stenting with the Svelte 'all-in-one' coronary stent Integrated Delivery System (IDS) through diagnostic catheters to identify the clinical indications for which this approach is appropriately suited. Methods & results: Forty-eight (48) patients with 54 lesions of lengths ≤20 mm and RVD 2.5-3.5 mm were targeted for direct stenting through diagnostic catheters (4-6F) via radial or femoral approach. Procedural characteristics early in an investigator's experience (28 lesions) were compared with outcomes following experience (26 lesions). Procedure, device and strategy success were realized in 54 (100%), 50 (93%) and 46 (85%) lesions, respectively, with strategy success significantly related to RVD (P = 0.05), lesion location (P = 0.01), and diagnostic catheter size (P = 0.05). Significant improvement in crossing and intervention time and trends toward improvement in device and strategy success, reductions in procedure and radiation time and contrast use were observed. Conclusions: Direct stenting through diagnostic catheters via radial or femoral approach using the Svelte IDS is feasible and associated with good in-hospital outcomes. This approach offers the attractive option of assessing lesions via diagnostic catheter and, depending upon vessel anatomy and lesion morphology, continuing with ad-hoc interventional treatment using the same diagnostic catheter. Improvements in strategy success and procedural efficiencies, based on operator experience, facilitate catheter downsizing and reduce intervention time, ancillary product use and overall procedure costs.

Keywords

Cardiac catheterization, Coronary artery disease diagnostic, Direct stenting, Percutaneous coronary intervention, Taverne, Cardiology and Cardiovascular Medicine, Radiology Nuclear Medicine and imaging, General Medicine, Journal Article, Multicenter Study

Citation

Khattab, A A, Nijhoff, F, Schofer, J, Berland, J, Meier, B, Nietlispach, F, Agostoni, P, Brucks, S & Stella, P 2015, 'Svelte Integrated Delivery System Performance Examined Through Diagnostic Catheter Delivery : The SPEED Registry', Catheterization and Cardiovascular Interventions, vol. 85, no. 1, pp. E23-E31. https://doi.org/10.1002/ccd.25621