Valve-in-Valve Implantation Using the ACURATE Neo in Degenerated Aortic Bioprostheses: An International Multicenter Analysis

Publication date

2019-11-25

Authors

Holzamer, Andreas
Kim, Won-Keun
Rück, Andreas
Sathananthan, Janarthanan
Keller, Lukas
Cosma, Joseph
Bauer, Timm
Nef, Holger
Amat-Santos, Ignacio J
Brinkert, Miriam

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

OBJECTIVES: This study reports an international experience using the transfemoral ACURATE neo transcatheter heart valve (Boston Scientific, Marlborough, Massachusetts) for the treatment of degenerated surgical aortic bioprostheses. BACKGROUND: Transcatheter valve-in-valve procedures have emerged as an alternative to redo surgery. Supra-annular prostheses might be particularly useful in this indication. METHODS: This is an international multicenter analysis including 85 patients from 14 centers in Europe and Canada undergoing an ACURATE neo valve-in-valve procedure from March 2015 to February 2019. RESULTS: Internal diameter of the degenerated bioprosthesis was 20.3 ± 2.1 mm. Prosthesis size S was used in 70 (82%) procedures. The median depth of implantation was 3 mm and the upper crown of the ACURATE neo was positioned above the stent posts of the degenerated bioprosthesis in 54 (64%) and inside in 31 (36%). Mean transvalvular gradient before discharge was significantly lower if the upper crown was above the degenerated bioprosthesis (13.7 ± 5.9 mm Hg vs. 19.5 ± 10.0 mm Hg; p = 0.001). However, a high position of the ACURATE neo resulted in embolization in 1 patient, conversion to open-heart surgery in 1, and need for reintervention due to transcatheter heart valve failure within the first 18 months of follow-up in 4. CONCLUSIONS: This early experience shows that a high implantation of the ACURATE neo with the upper crown above the stent posts of the degenerated bioprosthesis resulted in lower mean transvalvular gradients but a higher rate of malpositioning and early valve degeneration.

Keywords

AVA, aortic valve area, ID, inner diameter, PPM, prosthesis-patient mismatch, THV, transcatheter heart valve, ViV, valve-in-valve, transcatheter aortic valve implantation, transfemoral, transcatheter aortic valve replacement, valve-in-valve, degenerated bioprosthesis, Taverne, Cardiology and Cardiovascular Medicine, Journal Article

Citation

Holzamer, A, Kim, W-K, Rück, A, Sathananthan, J, Keller, L, Cosma, J, Bauer, T, Nef, H, Amat-Santos, I J, Brinkert, M, Husser, O, Pellegrini, C, Schofer, J, Nerla, R, Montorfano, M, Giannini, F, Stella, P, Kuwata, S, Hilker, M, Castriota, F, Ussia, G P, Webb, J G, Nietlispach, F & Toggweiler, S 2019, 'Valve-in-Valve Implantation Using the ACURATE Neo in Degenerated Aortic Bioprostheses : An International Multicenter Analysis', JACC. Cardiovascular Interventions, vol. 12, no. 22, pp. 2309-2316. https://doi.org/10.1016/j.jcin.2019.07.042