Abstract
Motion is problematic during radiotherapy as it could lead to potential underdosage of the tumor, and/or overdosage in organs-at-risk. A solution is adaptive radiotherapy guided by magnetic resonance imaging (MRI). MRI allows for imaging of target volumes and organs-at-risk before and during treatment delivery with superb soft tissue contrast in any desired orientation, enabling motion management by means of (real-time) adaptive radiotherapy. The noise navigator, which is independent of the MR signal, could serve as a secondary motion detection method in synergy with MR imaging. The feasibility of respiratory motion detection by means of the noise navigator was demonstrated previously. Furthermore, from electromagnetic simulations we know that the noise navigator is sensitive to tissue displacement and thus could in principle be used for the detection of various types of motion. In this study we demonstrate the detection of various types of motion for three anatomical use cases of MRI-guided radiotherapy, i.e. torso (bulk movement and variable breathing), head-and-neck (swallowing) and cardiac. Furthermore, it is shown that the noise navigator can detect bulk movement, variable breathing and swallowing on a hybrid 1.5 T MRI-linac system. Cardiac activity detection through the noise navigator seems feasible in an MRI-guided radiotherapy setting, but needs further optimization. The noise navigator is a versatile and fast (millisecond temporal resolution) motion detection method independent of MR signal that could serve as an independent verification method to detect the occurrence of motion in synergy with real-time MRI-guided radiotherapy.
Keywords
magnetic resonance imaging, noise navigator, physiological motion, radiotherapy, thermal noise, Taverne, Radiological and Ultrasound Technology, Radiology Nuclear Medicine and imaging, Journal Article
Citation
Navest, R J M, Mandija, S, Zijlema, S E, Stemkens, B, Andreychenko, A, Lagendijk, J J W & van den Berg, C A T 2020, 'The noise navigator for MRI-guided radiotherapy : an independent method to detect physiological motion', Physics in medicine and biology, vol. 65, no. 12, 12NT01. https://doi.org/10.1088/1361-6560/ab8cd8