Embrella embolic deflection device for cerebral protection during transcatheter aortic valve replacement

Publication date

2015-01-01

Authors

Samim, Mariam
Agostoni, PierfrancescoISNI 0000000388177910
Hendrikse, JeroenISNI 0000000390964171
Budde, Ricardo P JISNI 0000000392898112
Nijhoff, F
Kluin, Jolanda
Ramjankhan, Faiz Z.ISNI 0000000395925702
Doevendans, PieterISNI 0000000110574516
Stella, Pieter R.ISNI 0000000393574363

Editors

Advisors

Supervisors

Document Type

Article

Collections

License

Abstract

AIMS: To compare the extent of cerebral ischemic injury after transcatheter aortic valve replacement (TAVR) with the use of an Embrella Embolic Deflector System versus unprotected TAVR. METHODS: Fifteen patients with severe symptomatic aortic stenosis underwent TAVR with use of the Embrella Embolic Deflector System for cerebral protection. Cerebral diffusion-weighted magnetic resonance imaging (DWI) was performed in all patients at day 4 after the procedure and images were retrospectively compared to 37 patients who had previously undergone TAVR without a protection device (TAVR-only group). RESULTS: Successful placement of the Embrella device was achieved in all patients. DWI revealed an increase in the number of ischemic lesions in the Embrella group compared with the TAVR-only group (9.0 vs 5.0, P = .044). The use of the Embrella device was however associated with a significant reduction in single-lesion volume: 9.7 μL [5.8, 18.4] versus 17.8 μL [9.5, 38.7] (P < .001). Moreover, total infarct volumes of more than 1000 μL were only seen in the TAVR-only group. More lesions occurred in the right side of the brain in the Embrella group, whereas in the TAVR-only group lesions were distributed equally between left and right. One patient in the TAVR-only group suffered from a transient ischemic attack. Postoperative evaluation was clinically uneventful in the Embrella group. CONCLUSIONS: The use of the Embrella device during TAVR increased the number of cerebral ischemic lesions on postprocedural brain imaging. This increase in number was however accompanied by a significant reduction in single-lesion volume and the absence of large total infarct volumes.

Keywords

aortic valve stenosis, cerebral embolic protection device, cerebral ischemic injury, Embrella Embolic Deflector System, intracranial embolism, transcatheter aortic valve replacement, Cardiology and Cardiovascular Medicine, Surgery, Pulmonary and Respiratory Medicine, Comparative Study, Journal Article

Citation

Samim, M, Agostoni, P, Hendrikse, J, Budde, R P J, Nijhoff, F, Kluin, J, Ramjankhan, F, Doevendans, P A & Stella, P R 2015, 'Embrella embolic deflection device for cerebral protection during transcatheter aortic valve replacement', Journal of Thoracic and Cardiovascular Surgery, vol. 149, no. 3, pp. 799-805.e2. https://doi.org/10.1016/j.jtcvs.2014.05.097