A reusable 3D printed brain-like phantom for benchmarking electrical properties tomography reconstructions

Publication date

2024-11

Authors

Meerbothe, Thierry
Florczak, Sammy
van den Berg, CATORCID 0000-0002-5565-6889
Levato, RiccardoORCID 0000-0002-3795-3804
Mandija, StefanoORCID 0000-0002-4612-5509

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

PURPOSE: In MR electrical properties tomography (MR-EPT), electrical properties (EPs, conductivity and permittivity) are reconstructed from MR measurements. Phantom measurements are important to characterize the performance of MR-EPT reconstruction methods, since they allow knowledge of reference EPs values. To assess reconstruction methods in a more realistic scenario, it is important to test the methods using phantoms with realistic shapes, internal structures, and dielectric properties. In this work, we present a 3D printing procedure for the creation of realistic brain-like phantoms to benchmark MR-EPT reconstructions. METHODS: We created two brain-like geometries with three different compartments using 3D printing. The first geometry was filled once, while the second geometry was filled three times with different saline-gelatin solutions, resulting in a total of four phantoms with different EPs. The saline solutions were characterized using a probe. 3D MR-EPT reconstructions were performed from MR measurements at 3T. The reconstructed conductivity values were compared to reference values of the saline-gelatin solutions. The measured fields were also compared to simulated fields using the same phantom geometry and electrical properties. RESULTS: The measured fields were consistent with simulated fields. Reconstructed conductivity values were consistent with the reference (probe) conductivity values. This indicated the suitability of such phantoms for benchmarking MR-EPT reconstructions. CONCLUSION: We presented a new workflow to 3D print realistic brain-like phantoms in an easy and affordable way. These phantoms are suitable to benchmark MR-EPT reconstructions, but can also be used for benchmarking other quantitative MR methods.

Keywords

brain models, conductivity, electrical properties tomography, MR phantoms, MR-EPT, Radiology Nuclear Medicine and imaging, Journal Article

Citation

Meerbothe, T G, Florczak, S, van den Berg, C A T, Levato, R & Mandija, S 2024, 'A reusable 3D printed brain-like phantom for benchmarking electrical properties tomography reconstructions', Magnetic Resonance in Medicine, vol. 92, no. 5, pp. 2271-2279. https://doi.org/10.1002/mrm.30189