Identifying the source of spurious signals caused by B 0 inhomogeneities in single-voxel 1H MRS

Publication date

2022-07

Authors

Shams, Z
Klomp, Dennis W.J.ORCID 0000-0002-5884-5386ISNI 0000000396514008
Boer, Vincent O
Wijnen, JannieORCID 0000-0003-2352-3667ISNI 0000000387155509
Wiegers, Evita C.

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by_nc_nd

Abstract

PURPOSE: Single-voxel MRS (SV MRS) requires robust volume localization as well as optimized crusher and phase-cycling schemes to reduce artifacts arising from signal outside the volume of interest. However, due to local magnetic field gradients (B 0 inhomogeneities), signal that was dephased by the crusher gradients during acquisition might rephase, leading to artifacts in the spectrum. Here, we analyzed this mechanism, aiming to identify the source of signals arising from unwanted coherence pathways (spurious signals) in SV MRS from a B 0 map. METHODS: We investigated all possible coherence pathways associated with imperfect localization in a semi-localized by adiabatic selective refocusing (semi-LASER) sequence for potential rephasing of signals arising from unwanted coherence pathways by a local magnetic field gradient. We searched for locations in the B 0 map where the signal dephasing due to external (crusher) and internal (B 0 ) field gradients canceled out. To confirm the mechanism, SV-MR spectra (TE = 31 ms) and 3D-CSI data with the same volume localization as the SV experiments were acquired from a phantom and 2 healthy volunteers. RESULTS: Our analysis revealed that potential sources of spurious signals were scattered over multiple locations throughout the brain. This was confirmed by 3D-CSI data. Moreover, we showed that the number of potential locations where spurious signals could originate from monotonically decreases with crusher strength. CONCLUSION: We proposed a method to identify the source of spurious signals in SV 1 H MRS using a B 0 map. This can facilitate MRS sequence design to be less sensitive to experimental artifacts.

Keywords

Artifacts, Brain/diagnostic imaging, Humans, Magnetic Fields, Magnetic Resonance Imaging/methods, Phantoms, Imaging, H MRS, B inhomogeneity gradients, spurious signals, coherence pathways, Radiology Nuclear Medicine and imaging, Research Support, Non-U.S. Gov't, Journal Article

Citation

Shams, Z, Klomp, D W J, Boer, V O, Wijnen, J P & Wiegers, E C 2022, 'Identifying the source of spurious signals caused by B 0 inhomogeneities in single-voxel 1 H MRS', Magnetic Resonance in Medicine, vol. 88, no. 1, pp. 71-82. https://doi.org/10.1002/mrm.29222