Identifying patients at risk: multi-centre comparison of HeartMate 3 and HeartWare left ventricular assist devices
Publication date
2023-06
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by_nc
Abstract
AIMS: Since the withdrawal of HeartWare (HVAD) from the global market, there is an ongoing discussion if and which patients require prophylactically exchange for a HeartMate 3 (HM3). Therefore, it is important to study outcome differences between HVAD and HM3 patients. Because centres differ in patient selection and standard of care, we performed a propensity score (PS)-based study including centres that implanted both devices and aimed to identify which HVAD patients are at highest risk. METHODS AND RESULTS: We performed an international multi-centre study (n = 1021) including centres that implanted HVAD and HM3. PS-matching was performed using clinical variables and the implanting centre. Survival and complications were compared. As a sensitivity analysis, PS-adjusted Cox regression was performed. Landmark analysis with conditional survival >2 years was conducted to evaluate long-term survival differences. To identify which HVAD patients may benefit from a HM3 upgrade, Cox regression using pre-operative variables and their interaction with device type was performed. Survival was significantly better for HM3 patients (P < 0.01) in 458 matched patients, with a median follow-up of 23 months. Within the matched cohort, HM3 patients had a median age of 58 years, and 83% were male, 80% of the HVAD patients were male, with a median age of 59 years. PS-adjusted Cox regression confirmed a significantly better survival for HM3 patients when compared with HVAD, with a HR of 1.46 (95% confidence interval 1.14-1.85, P < 0.01). Pump thrombosis (P < 0.01) and ischaemic stroke (P < 0.01) occurred less in HM3 patients. No difference was found for haemorrhagic stroke, right heart failure, driveline infection, and major bleeding. Landmark-analysis confirmed a significant difference in conditional survival >2 years after implantation (P = 0.03). None of the pre-operative variable interactions in the Cox regression were significant. CONCLUSIONS: HM3 patients have a significantly better survival and a lower incidence of ischaemic strokes and pump thrombosis than HVAD patients. This survival difference persisted after 2 years of implantation. Additional research using post-operative variables is warranted to identify which HVAD patients need an upgrade to HM3 or expedited transplantation.
Keywords
Centrifugal continuous flow pump, End-stage heart failure, LVAD, Left ventricular assist device, Mechanical circulatory support, Cardiology and Cardiovascular Medicine
Citation
Numan, L, Zimpfer, D, Zadok, O I B, Aarts, E, Morshuis, M, Guenther, S P W, Riebandt, J, Wiedemann, D, Ramjankhan, F Z, Oppelaar, A-M, Ben-Gal, T, Ben-Avraham, B, Asselbergs, F W, Schramm, R & Van Laake, L W 2023, 'Identifying patients at risk : multi-centre comparison of HeartMate 3 and HeartWare left ventricular assist devices', ESC heart failure, vol. 10, no. 3, pp. 1656-1665. https://doi.org/10.1002/ehf2.14308