Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil
Publication date
2022-08-31
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by
Abstract
OBJECTIVES: Acoustic noise in magnetic resonance imaging (MRI) negatively impacts patients. We assessed a silent gradient coil switched at 20 kHz combined with a T 1-weighted magnetisation prepared rapid gradient-echo (MPRAGE) sequence at 7 T. METHODS: Five healthy subjects (21-29 years; three females) without previous 7-T MRI experience underwent both a quiet MPRAGE (Q-MPRAGE) and conventional MPRAGE (C-MPRAGE) sequence twice. Image quality was assessed quantitatively, and qualitatively by two neuroradiologists. Sound level was measured objectively and rated subjectively on a 0 to 10 scale by all subjects immediately following each sequence and after the whole examination (delayed). All subjects also reported comfort level, overall experience and willingness to undergo the sequence again. RESULTS: Compared to C-MPRAGE, Q-MPRAGE showed higher signal-to-noise ratio (10%; p = 0.012) and lower contrast-to-noise ratio (20%; p < 0.001) as well as acceptable to good image quality. Q-MPRAGE produced 27 dB lower sound level (76 versus 103 dB). Subjects reported lower sound level for Q-MPRAGE both immediate (4.4 ± 1.4 versus 6.4 ± 1.3; p = 0.007) and delayed (4.6 ± 1.4 versus 6.3 ± 1.3; p = 0.005), while they rated comfort level (7.4 ± 1.0 versus 6.1 ± 1.7; p = 0.016) and overall experience (7.6 ± 1.0 versus 6.0 ± 0.9; p = 0.005) higher. Willingness to undergo the sequence again was also higher, however not significantly (8.1 ± 1.0 versus 7.2 ± 1.3; p = 0.066). CONCLUSION: Q-MPRAGE using a silent gradient coil reduced sound level by 27 dB compared to C-MPRAGE at 7 T while featuring acceptable-to-good image quality and a quieter and more pleasant subject experience.
Keywords
Acoustics, Healthy volunteers, Magnetic resonance imaging, Neuroimaging, Noise, Brain/diagnostic imaging, Humans, Signal-To-Noise Ratio, Female, Magnetic Resonance Imaging/methods, Radiology Nuclear Medicine and imaging, Journal Article
Citation
Jacobs, S M, Versteeg, E, van der Kolk, A G, Visser, L N C, Oliveira, Í A F, van Maren, E, Klomp, D W J & Siero, J C W 2022, 'Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil', European radiology experimental, vol. 6, no. 1, 36, pp. 1-9. https://doi.org/10.1186/s41747-022-00293-x