Validation of transferability of DBA derivatization and LC-MS/MS determination method for isocyanates via an interlaboratory comparison.

Publication date

2008-11

Authors

Bobeldijk, I.
Karlsson, D.
Lai, Chen-Li
Gonsalves, J.
Hekman, M.
van de Lagemaat, D.
Preller, L.
Heederik, DickISNI 0000000388327640
Skarping, G.

Editors

Advisors

Supervisors

Document Type

Article
Open Access logo

License

Abstract

An adapted method for the quantitative determination of isocyanates in air was implemented and validated in-house. The method was based on air sampling using an impinger flask containing di-n-butylamine (DBA) in toluene and a glass fibre filter in series. The DBA derivatives were determined using liquid chromatography and tandem mass spectrometry. Studied isocyanates were isophorone diisocyanate, isocyanic acid (ICA), methyl isocyanate, ethyl isocyanate, propyl isocyanate, hexamethylene diisocyanate (HDI), 2,6- and 2,4-toluene diisocyanate, 4,4'-methylene diphenyl diisocyanate (MDI), phenyl isocyanate (PhI), MDI oligomers and different HDI adducts. Monitoring of selected reactions resulted in quantifications with correlation coefficients >0.995, within-batch relative standard deviation (RSD) of repeatability was 500 samples.

Keywords

interlaboratory method comparison, isocyanates, liquid chromatography mass spectrometry, DBA, Coronacrisis-Taverne

Citation

Bobeldijk, I, Karlsson, D, Pronk, A, Gonsalves, J, Hekman, M, van de Lagemaat, D, Preller, L, Heederik, D & Skarping, G 2008, 'Validation of transferability of DBA derivatization and LC-MS/MS determination method for isocyanates via an interlaboratory comparison.', Annals of Occupational Hygiene, vol. 52, no. 8, pp. 757-763. https://doi.org/10.1093/annhyg/men057