Challenges in Removing the Essure® Device
Publication date
2020
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
Abstract
We present a case about Essure® removal surgery in which the third markers of the device have torn off. The woman needed a second surgery for complete removal of the devices. Fluoroscopy during surgery is a good method to visualize the lost fragments. With fluoroscopy, a hysterectomy is not needed for complete removal. It is important to understand the structure of the device and to be aware of the four radiopaque markers during surgery and their removal.
Keywords
Case Reports
Citation