Calibration of the photon spectrometer PHOS of the ALICE experiment

Publication date

2019-05-28

Authors

ALICE Collaboration

Editors

Advisors

Supervisors

Document Type

Article
Open Access logo

License

Abstract

The procedure for the energy calibration of the high granularity electromagnetic calorimeter PHOS of the ALICE experiment is presented. The methods used to perform the relative gain calibration, to evaluate the geometrical alignment and the corresponding correction of the absolute energy scale, to obtain the nonlinearity correction coefficients and finally, to calculate the time-dependent calibration corrections, are discussed and illustrated by the PHOS performance in proton-proton (pp) collisions at √s=13 TeV. After applying all corrections, the achieved mass resolutions for π0 and η mesons for pT > 1.7 GeV/c are σm π0 = 4.56 ± 0.03 MeV/c2 and σm η = 15.3 ± 1.0 MeV/c2, respectively.

Keywords

Calorimeters, Detector alignment and calibration methods (lasers, sources, particlebeams), Mathematical Physics, Instrumentation

Citation

ALICE Collaboration 2019, 'Calibration of the photon spectrometer PHOS of the ALICE experiment', Journal of Instrumentation, vol. 14, no. 5, P05025. https://doi.org/10.1088/1748-0221/14/05/P05025