System for Image-Guided Resection of Non-Palpable Breast Lesions: Proof of Concept

Publication date

2018-05

Authors

Arsenali, B.
de Jong, Hugo W A MORCID 0000-0002-3000-8316
Viergever, MaxORCID 0000-0003-2582-042XISNI 0000000117491940
Gilhuijs, KennethORCID 0000-0003-2087-8649ISNI 0000000393336330

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

Abstract

Purpose: In breast-conserving surgery (BCS), the cancer is sometimes incompletely excised, leading to reduced patient survival. To pursue complete excisions, radioactive seed localization (RSL) may be used to insert an iodine-125 seed into the tumor. The seed is used as a marker for the location of the tumor during surgery. RSL does not, however, show the extent of the tumor. Based on RSL, we pursue to visualize the seed location together with the extent from diagnostic images. Methods: A system with two gamma-camera heads and two parallel-hole collimators was recently proposed to triangulate the location of an iodine-125 seed during BCS. In the present study, this system was extended with a range camera to visualize a sphere centered on the seed in relation to the breast. This sphere contains the entire tumor and thus defines the target volume for BCS. Physical experiments with acrylic block phantoms (thickness ranging from 3.5 to 6.5 cm) were performed to assess the absolute bias and the precision with which this sphere can be visualized. Results: When a 6.5 cm thick phantom was used, along the horizontal plane, the target volume was visualized with an absolute bias and a precision of 2.1 and 0.8 mm, respectively. Along the vertical axis (i.e., z-axis), these values were 4.2 and 2.8 mm, respectively. Conclusions: The proposed system visualizes the target volume with an absolute bias that may be acceptable for BCS.

Keywords

breast-conserving surgery, gamma-camera system, intraoperative guidance, radioactive seed localization, range-camera system, Biophysics, Radiology Nuclear Medicine and imaging

Citation

Arsenali, B, de Jong, H W A M, Viergever, M A & Gilhuijs, K G A 2018, 'System for Image-Guided Resection of Non-Palpable Breast Lesions : Proof of Concept', Medical Physics, vol. 45, no. 5, pp. 2169-2178. https://doi.org/10.1002/mp.12884