Off-Resonance Suppression for Multispectral MR Imaging Near Metallic Implants

Publication date

2015-01

Authors

den Harder, J. Chiel
van Yperen, Gert H.
Blume, Ulrike A.
Bos, CORCID 0000-0002-9246-3242ISNI 0000000388845122

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

Purpose: Metal artifact reduction in MRI within clinically feasible scan-times without through-plane aliasing. Theory and Methods: Existing metal artifact reduction techniques include view angle tilting (VAT), which resolves in-plane distortions, and multispectral imaging (MSI) techniques, such as slice encoding for metal artifact correction (SEMAC) and multi-Acquisition with variable resonances image combination (MAVRIC), that further reduce image distortions, but significantly increase scan-time. Scan-time depends on anatomy size and anticipated total spectral content of the signal. Signals outside the anticipated spatial region may cause through-plane back-folding. Off-resonance suppression (ORS), using different gradient amplitudes for excitation and refocusing, is proposed to provide well-defined spatial-spectral selectivity in MSI to allow scan-time reduction and flexibility of scan-orientation. Comparisons of MSI techniques with and without ORS were made in phantom and volunteer experiments. Results: Off-resonance suppressed SEMAC (ORS-SEMAC) and outer-region suppressed MAVRIC (ORS-MAVRIC) required limited through-plane phase encoding steps compared with original MSI. Whereas SEMAC (scan time: 5'46") and MAVRIC (4'12") suffered from through-plane aliasing, ORS-SEMAC and ORS-MAVRIC allowed alias-free imaging in the same scantimes. Conclusion: ORS can be used in MSI to limit the selected spatial-spectral region and contribute to metal artifact reduction in clinically feasible scan-times while avoiding slice aliasing.

Keywords

magnetic resonance imaging, arthroplasty, SEMAC, MAVRIC, metal implants, multispectral imaging, ARTIFACT CORRECTION, SPIN-ECHO, SUSCEPTIBILITY, ACQUISITION, REDUCTION, Taverne, Evaluation Studies, Journal Article

Citation

den Harder, J C, van Yperen, G H, Blume, U A & Bos, C 2015, 'Off-Resonance Suppression for Multispectral MR Imaging Near Metallic Implants', Magnetic Resonance in Medicine, vol. 73, no. 1, pp. 233-243. https://doi.org/10.1002/mrm.25126