Arterial pressure waveform analysis versus thermodilution cardiac output measurement during open abdominal aortic aneurysm repair: A prospective observational study

Publication date

2015-01-01

Authors

Montenij, Leon JISNI 0000000393466573
Buhre, Wolfgang F.
De Jong, Steven A.
Harms, Jeroen H.
van Herwaarden, JoostORCID 0000-0003-1165-5179ISNI 0000000393686613
Kruitwagen, Cas L J JISNI 0000000393886033
De Waal, Eric E CISNI 0000000388493008

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

BACKGROUND Arterial pressure waveform analysis enables continuous, minimally invasive measurement of cardiac output. Haemodynamic instability compromises the reliability of the technique and a means of maintaining accurate measurement in this circumstance would be useful. OBJECTIVES To investigate the accuracy, precision and trending ability of arterial pressure waveform cardiac output obtained with FloTrac/Vigileo, versus pulmonary artery thermodilution in patients undergoing elective open abdominal aortic aneurysm repair. DESIGN A prospective observational study. SETTING Operating room in a university hospital. PATIENTS Twenty-two patients scheduled for elective, open abdominal aortic aneurysm repair. MAIN OUTCOME MEASURES Bias, limits of agreement and mean error as determined with Bland-Altman analysis between arterial waveform and thermodilution cardiac output assessment at four time points: after induction of anaesthesia (t<inf>1</inf>); after aortic cross-clamping (t<inf>2</inf>); after clamp release (t<inf>3</inf>); and after skin closure (t<inf>4</inf>). Trending ability from t1 to t<inf>2</inf>, t<inf>2</inf> to t<inf>3</inf> and t<inf>3</inf> to t<inf>4</inf>, determined with four-quadrant and polar plot methodology. Clinically acceptable boundaries were defined in advance. RESULTS Bland-Altman analysis revealed a bias of 0.54 l min<sup>-1</sup> (thermodilution minus arterial waveform cardiac output) for pooled data, and 0.51 (t<inf>1</inf>), -0.42 (t<inf>2</inf>), 0.98 (t<inf>3</inf>) and 0.98 (t<inf>4</inf>) l min<sup>-1</sup> at the different time points. Limits of agreement (LOA) were [-3.0 to 4.0] (pooled), [-2.0 to 3.0] (t<inf>1</inf>), [-3.1 to 2.3] (t<inf>2</inf>), [-2.5 to 4.4] (t<inf>3</inf>) and [<sup>-1</sup>.7 to 3.7] (t<inf>4</inf>) l min<sup>-1</sup>, resulting in mean errors of 58% (pooled), 45% (t<inf>1</inf>), 53% (t<inf>2</inf>), 52% (t<inf>3</inf>) and 41% (t<inf>4</inf>). Four-quadrant concordance was 65%. Polar plot analysis resulted in an angular bias of <sup>-1</sup>28, with radial LOA of -60° to 36°. CONCLUSION Bias between arterial waveform and thermodilution cardiac output was within a predefined acceptable range, but the mean error was above the accepted range of 30%. Trending ability was poor. Arterial waveform and thermodilution cardiac outputs are, therefore, not interchangeable in patients undergoing open abdominal aortic aneurysm repair.

Keywords

Taverne, Anesthesiology and Pain Medicine, General Medicine, Comparative Study, Journal Article, Observational Study, Research Support, Non-U.S. Gov't

Citation

Montenij, L J, Buhre, W F, De Jong, S A, Harms, J H, Van Herwaarden, J A, Kruitwagen, C L J J & De Waal, E E C 2015, 'Arterial pressure waveform analysis versus thermodilution cardiac output measurement during open abdominal aortic aneurysm repair : A prospective observational study', European Journal of Anaesthesiology, vol. 32, no. 1, pp. 13-19. https://doi.org/10.1097/EJA.0000000000000160