Portable fixed dynamometry: towards remote muscle strength measurements in patients with motor neuron disease

Publication date

2021-05

Authors

Bakers, Jaap N EORCID 0000-0001-9881-7710
van den Berg, Leonard H.ISNI 0000000388137302
Ajeks, Toju G
Holleman, Maxine J
Verhoeven, Jill
Beelen, AnitaORCID 0000-0002-1269-0710
Visser-Meily, J. M.A.ISNI 0000000387554577
Van Eijk, RubenORCID 0000-0002-7132-5967

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

BACKGROUND: We aimed to determine (1) the test-retest reliability of a newly developed portable fixed dynamometer (PFD) as compared to the hand-held dynamometer (HHD) in patients with motor neuron disease (MND) and (2) the PFD's ability to reduce possible examiner-induced ceiling effects. METHODS: Test-retest reliability of isometric muscle strength of the quadriceps was measured in patients with MND and non-neurological controls using the HHD and PFD. Reliability was estimated by the intraclass correlation coefficient (ICC) and standard error of measurement (SEM) using linear mixed effects models, and the Bland-Altman method of agreement. RESULTS: In total, 45 patients with MND and 43 healthy controls were enrolled in this study. The ICC of the PFD was excellent and similar in both patients and controls (ICC Patients 99.5% vs. ICC Controls 98.6%) with a SEM of 6.2%. A strong examiner-induced ceiling effect in HHD was found when the participant's strength exceeded that of examiner. Employing the PFD increased the range of muscle strength measurements across individuals nearly twofold from 414 to 783 N. CONCLUSIONS: Portable fixed dynamometry may significantly reduce examiner-induced ceiling effects, optimize the standardization of muscle strength testing, and maximize reliability. Ultimately, PFD may improve the delivery of care due to its potential for unsupervised, home-based assessments and reduce the burden to the patient of participating in clinical trials for MND or other neuromuscular diseases.

Keywords

Humans, Isometric Contraction, Motor Neuron Disease, Muscle Strength, Muscle Strength Dynamometer, Reproducibility of Results, Muscle strength, Fixed dynamometry, Motor neuron disease, Reliability study, Remote monitoring, Clinical Neurology, Neurology, Journal Article

Citation

Bakers, J N E, van den Berg, L H, Ajeks, T G, Holleman, M J, Verhoeven, J, Beelen, A, Visser-Meily, J M A & van Eijk, R P A 2021, 'Portable fixed dynamometry : towards remote muscle strength measurements in patients with motor neuron disease', Journal of Neurology, vol. 268, no. 5, pp. 1738-1746. https://doi.org/10.1007/s00415-020-10366-9