Exercise MRI stress testing of the human heart at 3 Tesla: measurement precision of biventricular function and aortic blood flow during steady-state bicycling exercise

Publication date

2026-04

Authors

Klarenberg, Hugo
Froeling, MartijnORCID 0000-0003-3841-0497
Leiner, TimORCID 0000-0003-1885-5499ISNI 0000000390698205
Lamb, Hildo J
Boekholdt, S Matthijs
Jørstad, Harald T
Strijkers, Gustav J
Bakermans, Adrianus J

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

OBJECTIVE: This work aimed to demonstrate the feasibility of quantifying heart function during bicycling exercise with dynamic real-time cine MRI at 3 Tesla, and to assess its measurement precision. MATERIALS AND METHODS: Twelve volunteers performed steady-state bicycling exercise, while real-time cine MR images were collected using a 72-channel receiver coil array and a parallel imaging acceleration factor of 5. Biventricular end-diastolic and end-systolic (ESV) volumes and function during exercise were compared with resting-state real-time cine MRI and conventional cardiac-gated cine MRI under breath holding, and validated against 2D phase-contrast MRI-based estimates of aortic blood flow. Precision was evaluated as the inter-session measurement repeatability. RESULTS: Left (LV) and right ventricular (RV) stroke volumes (SV) increased progressively with exercise intensity, which was mediated by a decrease in ESV. Likewise, LV SV estimated with 2D phase-contrast MRI increased from 90 ± 17 mL at rest to 114 ± 29 mL during vigorous-intensity exercise. Repeatability coefficients were 52% and 41% for LV SV at moderate- and vigorous-intensity exercise, while RV SV repeatability coefficients were 58% and 42%, respectively. DISCUSSION: We established an exercise MRI stress testing protocol for quantifying biventricular volumes and function during moderate- and vigorous-intensity steady-state bicycling exercise.

Keywords

Accelerated MRI, Athlete’s heart, Ergometry, Exercise intolerance, HFpEF, Biophysics, Radiological and Ultrasound Technology, Radiology Nuclear Medicine and imaging

Citation

Klarenberg, H, Froeling, M, Leiner, T, Lamb, H J, Boekholdt, S M, Jørstad, H T, Strijkers, G J & Bakermans, A J 2026, 'Exercise MRI stress testing of the human heart at 3 Tesla : measurement precision of biventricular function and aortic blood flow during steady-state bicycling exercise', Magma - Magnetic Resonance Materials In Physics Biology And Medicine, vol. 39, no. 2, pp. 187-200. https://doi.org/10.1007/s10334-025-01304-9