Interference suppression for EEG recording during open heart surgery

Publication date

1979-05

Authors

Weide, H. van der
Pronk, R.A.F.

Editors

Advisors

Supervisors

DOI

Document Type

Article
Open Access logo

License

Abstract

A device for recording the EEG during open heart surgery is described. It differs from most standard equipment in two ways. First, the input circuit is completely floating from earth and will withstand 500 V DC. Second, radiofrequency (RF)_shielding and filtering permits continuous recording of the EEG during electro-surgery in the operating room. The almost undisturbed recording permits on-line continuous EEG monitoring by means of quantitative signal analysis methods.