Intrinsic variability of fluorescence calibrators impacts the assignment of MESF or ERF values to nanoparticles and extracellular vesicles by flow cytometry

Publication date

2024-02

Authors

Lozano-Andrés, EstefaníaISNI 0000000492910342
Van Den Broeck, Tina
Wang, Lili
Mehrpouyan, Majid
Tian, Ye
Yan, Xiaomei
Arkesteijn, Ger J A
Wauben, MarcaORCID 0000-0003-0360-0311ISNI 0000000390143250

Editors

Advisors

Supervisors

Document Type

Article
Open Access logo

License

cc_by

Abstract

Flow cytometry allows to characterize nanoparticles (NPs) and extracellular vesicles (EVs) but results are often expressed in arbitrary units of fluorescence. We evaluated the precision and accuracy of molecules of equivalent soluble fluorophores (MESF) beads for calibration of NPs and EVs. Firstly, two FITC-MESF bead sets, 2 and 6 um in size, were measured on three flow cytometers. We showed that arbitrary units could not be compared between instruments but after calibration, comparable FITC MESF units were achieved. However, the two calibration bead sets displayed varying slopes that were consistent across platforms. Further investigation revealed that the intrinsic uncertainty related to the MESF beads impacts the robust assignment of values to NPs and EVs based on extrapolation into the dim fluorescence range. Similar variations were found with PE MESF calibration. Therefore, the same calibration materials and numbers of calibration points should be used for reliable comparison of submicron sized particles.

Keywords

Extracellular vesicles, Flow cytometry, Fluorescence calibration, Nanoparticles, Standardization

Citation

Lozano-Andrés, E, Van Den Broeck, T, Wang, L, Mehrpouyan, M, Tian, Y, Yan, X, Arkesteijn, G J A & Wauben, M H M 2024, 'Intrinsic variability of fluorescence calibrators impacts the assignment of MESF or ERF values to nanoparticles and extracellular vesicles by flow cytometry', Nanomedicine: Nanotechnology, Biology, and Medicine, vol. 56, 102720. https://doi.org/10.1016/j.nano.2023.102720