Intrinsic variability of fluorescence calibrators impacts the assignment of MESF or ERF values to nanoparticles and extracellular vesicles by flow cytometry
Publication date
2024-02
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by
Abstract
Flow cytometry allows to characterize nanoparticles (NPs) and extracellular vesicles (EVs) but results are often expressed in arbitrary units of fluorescence. We evaluated the precision and accuracy of molecules of equivalent soluble fluorophores (MESF) beads for calibration of NPs and EVs. Firstly, two FITC-MESF bead sets, 2 and 6 um in size, were measured on three flow cytometers. We showed that arbitrary units could not be compared between instruments but after calibration, comparable FITC MESF units were achieved. However, the two calibration bead sets displayed varying slopes that were consistent across platforms. Further investigation revealed that the intrinsic uncertainty related to the MESF beads impacts the robust assignment of values to NPs and EVs based on extrapolation into the dim fluorescence range. Similar variations were found with PE MESF calibration. Therefore, the same calibration materials and numbers of calibration points should be used for reliable comparison of submicron sized particles.
Keywords
Extracellular vesicles, Flow cytometry, Fluorescence calibration, Nanoparticles, Standardization
Citation
Lozano-Andrés, E, Van Den Broeck, T, Wang, L, Mehrpouyan, M, Tian, Y, Yan, X, Arkesteijn, G J A & Wauben, M H M 2024, 'Intrinsic variability of fluorescence calibrators impacts the assignment of MESF or ERF values to nanoparticles and extracellular vesicles by flow cytometry', Nanomedicine: Nanotechnology, Biology, and Medicine, vol. 56, 102720. https://doi.org/10.1016/j.nano.2023.102720