Performance analysis of a dedicated breast MR-HIFU system for tumor ablation in breast cancer patients

Publication date

2015-07-21

Authors

Deckers, R.
Merckel, L. G.
de Senneville, B. Denis
Schubert, G.
Koehler, M.
Knuttel, F.M.
Mali, WPTMISNI 0000000392849126
Moonen, C. T. W.ORCID 0000-0001-5593-3121ISNI 0000000038813649
van den Bosch, MauriceISNI 0000000387826245
Bartels, WilbertISNI 0000000388733745

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

taverne

Abstract

MR-guided HIFU ablation is a promising technique for the non-invasive treatment of breast cancer. A phase I study was performed to assess the safety and treatment accuracy and precision of MR-HIFU ablation in breast cancer patients (n = 10) using a newly developed MR-HIFU platform dedicated to applications in the breast. In this paper a technical analysis of the performance of the dedicated breast MR-HIFU system during breast tumors ablation is described. The main points of investigation were the spatial targeting accuracy and precision of the system and the performance of real-time respiration-corrected MR thermometry. The mean targeting accuracy was in the range of 2.4-2.6 mm, whereas the mean targeting precision was in the range of 1.5-1.8 mm. To correct for respiration-induced magnetic field fluctuations during MR temperature mapping a look-up-table (LUT)-based correction method was used. An optimized procedural sedation protocol in combination with the LUT-based correction method allowed for precise MR thermometry during the ablation procedure (temperature standard deviation

Keywords

high intensity focused ultrasound, thermometry, breast cancer, MRI, FOCUSED ULTRASOUND SURGERY, CONTRAST-ENHANCED MRI, 20-YEAR FOLLOW-UP, RADIOFREQUENCY ABLATION, THERMOMETRY, THERAPY, US, FEASIBILITY, MAMMOGRAPHY, TISSUE, Taverne, Comparative Study, Journal Article, Research Support, Non-U.S. Gov't

Citation

Deckers, R, Merckel, L G, de Senneville, B D, Schubert, G, Koehler, M, Knuttel, F M, Mali, W P T M, Moonen, C T W, van den Bosch, M & Bartels, L W 2015, 'Performance analysis of a dedicated breast MR-HIFU system for tumor ablation in breast cancer patients', Physics in Medicine and Biology, vol. 60, no. 14, pp. 5527-5542. https://doi.org/10.1088/0031-9155/60/14/5527