Physical basis of ballistocardiography. IV. The relative movement of subject and ballistocardiograph
Publication date
1956-11
Authors
Burger, H.C.
Noordergraaf, A.
Korsten, J.J.M.
Ullersma, P.
Editors
Advisors
Supervisors
DOI
Document Type
Article
Metadata
Show full item recordCollections
License
Abstract
Amplitude characteristics have been calculated numerically from formulas derived in this paper for the low-frequency bcg (Gordon, Henderson, Burger), the middle-frequency bcg (Nickerson), and the high-frequency bcg's (Starr and Dock), taking the difference in movement of subject and bcg into account. In a few cases corresponding phase characteristics have been calculated. Two constants of the system were determined experimentally.
From this it appears that it is only the low-frequency bcg that gives ballistocardiograms that are distorted slightly. Moreover, the velocity and the acceleration can be recorded with an analogously slight distortion.