Non-invasive blood pressure measurement in ferrets (Mustela putorius furo) using high definition oscillometry

Publication date

2017-10-01

Authors

van Zeeland, Yvonne R.A.ISNI 0000000419447363
Wilde, A.
Bosman, I.H.
Uilenreef, JoostISNI 0000000387736856
Egner, B.I.
Schoemaker, Nico J.ISNI 0000000395017501

Editors

Advisors

Supervisors

Document Type

Article
Open Access logo

License

Abstract

This study was conducted to validate the use of high definition oscillometry (HDO) for non-invasive blood pressure (NIBP) measurements in ferrets and to establish reference ranges for NIBP in minimally sedated, healthy, young adult ferrets (170 mmHg) conditions. Although HDO correlated well with DABP (r > 0.90), it showed significant proportional bias, whereby HDO generally underestimated DABP with hyper- and normotensive conditions, and overestimated DABP with hypotensive conditions. Measurements obtained from the hind limb showed higher bias than those obtained from the tail or forelimb (P <0.001). Based on the above, and for practical reasons, the tail was selected as the preferred site of cuff placement in ferrets. Subsequently, a cross-over study was performed in 10 ferrets to establish the minimum dose of butorphanol and midazolam needed to successfully obtain NIBP in 100% of cases. Using this dose (0.2 mg/kg IM, each), reference intervals for NIBP from 63 healthy, young adult ferrets were established at 95–155 mmHg (systolic), 69–109 mmHg (mean) and 51–87 mmHg (diastolic) arterial pressures.

Keywords

Ferrets, High definition oscillometry, Non-invasive blood pressure measurement, Reference ranges

Citation

van Zeeland, Y R A, Wilde, A, Bosman, I H, Uilenreef, J J, Egner, B I & Schoemaker, N J 2017, 'Non-invasive blood pressure measurement in ferrets (Mustela putorius furo) using high definition oscillometry', Veterinary Journal, vol. 228, pp. 53-62. https://doi.org/10.1016/j.tvjl.2017.10.019