Abstract
Therapeutic monoclonal antibodies (mAbs) are rapidly taking over the treatment of many malignancies, and an astonishing number of mAbs is in development. This causes a high demand for quantification of mAbs in biomatrices both for measuring therapeutic mAb concentrations and to support pharmacokinetics and pharmacodynamics studies. Conventionally, ligand-binding assays are used for these purposes, but LC-MS is gaining popularity. Although intact (top-down) and subunit (middle-down) mAb quantification is reported, signature peptide (bottom-up) quantification is currently most advantageous. This review provides an overview of the reported bottom-up mAb quantification methods in biomatrices as well as general recommendations regarding signature peptide and internal standard selection, reagent use and optimization of digestion in bottom-up quantification methods.
Keywords
Taverne, SDG 3 - Good Health and Well-being
Citation
de Jong, K A, van Breugel, S J, Hillebrand, M J, Rosing, H, Huitema, A D & Beijnen, J H 2020, 'Bottom-up sample preparation for the LC-MS/MS quantification of anti-cancer monoclonal antibodies in bio matrices', Bioanalysis, vol. 12, no. 19, pp. 1405-1425. https://doi.org/10.4155/bio-2020-0204