Effect of non-return valves on the time-of-arrival of new medication in a patient after syringe exchange in an infusion set-up
Publication date
2023-02-06
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
cc_by
Abstract
The presence of a non-return valve in an infusion set-up is expected to affect the time-of-arrival of new medication in a patient after syringe exchange. Using Computational Fluid Dynamics (CFD) we have studied the flow through a typical non-return valve, focusing on two separate effects: (A) the overall delay in the time-of-arrival, and (B) timing effects due to the distortion of the Poiseuille flow profile in the non-return valve. The results show that (A) the additional delay in time-of-arrival of new medication, caused by the non-return valve alone, corresponds to the delay that would be caused by 11.2 cm of extra infusion line instead of the valve, and that (B) the non-Poiseuille flow profile inside the non-return valve gives rise to an extra slow wash-out of the last portion of the remnant fluid of the old medication. We conclude that awareness of these extra delays may be important for clinicians in certain time-critical situations.
Keywords
CFD, dosing errors, drug delivery, infusion line, non-return valve, Poiseuille flow, CFD, infusion line, non-return valve, Poiseuille flow, drug delivery, dosing errors, General Medicine, Biomedical Engineering
Citation
Konings, MK, Gevers, R, Mejri, S & Timmerman, AMDE 2023, 'Effect of non-return valves on the time-of-arrival of new medication in a patient after syringe exchange in an infusion set-up', Biomedizinische Technik, vol. 68, no. 1, 8, pp. 91-96. https://doi.org/10.1515/bmt-2022-0054