Practical considerations of diffusion-weighted MRS with ultra-strong diffusion gradients

Publication date

2023-12-07

Authors

Davies-Jenkins, Christopher W
Döring, André
Fasano, Fabrizio
Kleban, Elena
Mueller, Lars
Evans, C John
Afzali, Maryam
Jones, Derek K
Ronen, Itamar
Branzoli, Francesca

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

cc_by

Abstract

INTRODUCTION: Diffusion-weighted magnetic resonance spectroscopy (DW-MRS) offers improved cellular specificity to microstructure-compared to water-based methods alone-but spatial resolution and SNR is severely reduced and slow-diffusing metabolites necessitate higher b-values to accurately characterize their diffusion properties. Ultra-strong gradients allow access to higher b-values per-unit time, higher SNR for a given b-value, and shorter diffusion times, but introduce additional challenges such as eddy-current artefacts, gradient non-uniformity, and mechanical vibrations. METHODS: In this work, we present initial DW-MRS data acquired on a 3T Siemens Connectom scanner equipped with ultra-strong (300 mT/m) gradients. We explore the practical issues associated with this manner of acquisition, the steps that may be taken to mitigate their impact on the data, and the potential benefits of ultra-strong gradients for DW-MRS. An in-house DW-PRESS sequence and data processing pipeline were developed to mitigate the impact of these confounds. The interaction of TE, b-value, and maximum gradient amplitude was investigated using simulations and pilot data, whereby maximum gradient amplitude was restricted. Furthermore, two DW-MRS voxels in grey and white matter were acquired using ultra-strong gradients and high b-values. RESULTS: Simulations suggest T2-based SNR gains that are experimentally confirmed. Ultra-strong gradient acquisitions exhibit similar artefact profiles to those of lower gradient amplitude, suggesting adequate performance of artefact mitigation strategies. Gradient field non-uniformity influenced ADC estimates by up to 4% when left uncorrected. ADC and Kurtosis estimates for tNAA, tCho, and tCr align with previously published literature. DISCUSSION: In conclusion, we successfully implemented acquisition and data processing strategies for ultra-strong gradient DW-MRS and results indicate that confounding effects of the strong gradient system can be ameliorated, while achieving shorter diffusion times and improved metabolite SNR.

Keywords

diffusion-weighted MRS, eddy currents, gradient non-uniformity, metabolites, ultra-strong gradients, Journal Article

Citation

Davies-Jenkins, C W, Döring, A, Fasano, F, Kleban, E, Mueller, L, Evans, C J, Afzali, M, Jones, D K, Ronen, I, Branzoli, F & Tax, C M W 2023, 'Practical considerations of diffusion-weighted MRS with ultra-strong diffusion gradients', Frontiers in Neuroscience, vol. 17, 1258408. https://doi.org/10.3389/fnins.2023.1258408