Personal exposure to static and time-varying magnetic fields during MRI system test procedures.

Files

Access status: Embargo until 2050-01-01 , Vocht-J_Magn_Reson_Imaging_2009_30_1223-1228.pdf (237.92 KB)

Publication date

2009

Authors

de Vocht, F.
Muller, F.
Engels, H.
Kromhout, HansORCID 0000-0002-4233-1890ISNI 0000000033136431

Editors

Advisors

Supervisors

DOI

Document Type

Article

License

Abstract

PURPOSE: To assess personal time-weighted average (TWA) static magnetic field and time-varying magnetic field exposure for several system testing tasks routinely conducted by engineers at an MRI manufacturing plant. MATERIALS AND METHODS: Personal exposure was measured using a personal dosimeter that measured B and dB/dt in all three orthogonal directions with a 1 Hz frequency. TWA exposure was calculated and random effects and linear mixed effects models were used to assess exposure levels and variability. RESULTS: Whereas full-body peak (2T) and TWA (200 mT/8 h) exposure limits for the static magnetic field originally proposed by ICNIRP were not exceeded, time-varying magnetic field peaks (dB/dt) did exceed the proposed threshold value (767.9 mT/s) suggested by IEEE on several occasions at the 3.0 Tesla (T) whole-body cylindrical system. CONCLUSION: TWA exposure levels are well below the proposed limit values, but peak exposure limits are exceeded during certain procedures performed on 1.0T to 3.0T systems.

Keywords

Citation

de Vocht, F, Muller, F, Engels, H & Kromhout, H 2009, 'Personal exposure to static and time-varying magnetic fields during MRI system test procedures.', Journal of Magnetic Resonance Imaging, vol. 30, no. 5, pp. 1223-1228.