MRI-based transfer function determination for the assessment of implant safety
Files
Publication date
2017-12
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
taverne
Abstract
Purpose: We introduce a new MR-based method to determine the transfer function (TF) for radiofrequency (RF) safety assessment of active implantable medical devices. Transfer functions are implant-specific measures that relate the incident tangential electric field on an (elongated) implant to a scattered electric field at its tip. The proposed method allows for TF determination with a high spatial resolution in relatively fast measurements without requiring dedicated bench setups from MRI images. Theory and Methods: The principle of reciprocity is used in conjunction with the potential to measure currents with MRI to determine TF. Low-flip angle 3D dual gradient echo MRI data are acquired with an implant as transceive antenna, which requires minimal hardware adaptations. The implant-specific TF is determined from the acquired MRI data, with two different postprocessing methods for comparison. Results: TFs of linear and helical implants can be determined accurately (with a Pearson correlation coefficient R ≥ 0.7 between measurements and simulations, and a difference in field at the tip ΔEtip ≤ 19%) from relatively quick (t < 20 minutes) MRI acquisitions with (several) millimeter spatial resolution. Conclusion: Transfer function determination with MRI for RF safety assessment of implantable medical devices is possible. The proposed MR-based method allows for TF determination in more realistic exposure scenarios and solid media. Magn Reson Med 78:2449–2459, 2017.
Keywords
EM simulations, RF heating, active implantable medical device (AIMD), safety, transfer function, Taverne, Journal Article
Citation
Tokaya, J P, Raaijmakers, A J E, Luijten, P R, Bakker, J F & van den Berg, C A T 2017, 'MRI-based transfer function determination for the assessment of implant safety', Magnetic Resonance in Medicine, vol. 78, no. 6, pp. 2449-2459. https://doi.org/10.1002/mrm.26613