Acceptance procedure for the linear accelerator component of the 1.5 T MRI-linac

Publication date

2021-08

Authors

Woodings, Simon J
de Vries, J. H.Wilfred
Kok, J. G MISNI 0000000391574140
Hackett, Sara L
van Asselen, BISNI 0000000391220676
Bluemink, Johanna J.
Van Zijp, H. M.
van Soest, Theo L.
Roberts, David A.
Lagendijk, J J WISNI 0000000393637862

Editors

Advisors

Supervisors

Document Type

Article

Collections

Open Access logo

License

Abstract

Purpose: To develop and implement an acceptance procedure for the new Elekta Unity 1.5 T MRI-linac. Methods: Tests were adopted and, where necessary adapted, from AAPM TG106 and TG142, IEC 60976 and NCS 9 and NCS 22 guidelines. Adaptations were necessary because of the atypical maximum field size (57.4 × 22 cm), FFF beam, the non-rotating collimator, the absence of a light field, the presence of the 1.5 T magnetic field, restricted access to equipment within the bore, fixed vertical and lateral table position, and the need for MR image to MV treatment alignment. The performance specifications were set for stereotactic body radiotherapy (SBRT). Results: The new procedure was performed similarly to that of a conventional kilovoltage x-ray (kV) image guided radiation therapy (IGRT) linac. Results were acquired for the first Unity system. Conclusions: A comprehensive set of tests was developed, described and implemented for the MRI-linac. The MRI-linac met safety requirements for patients and operators. The system delivered radiation very accurately with, for example a gantry rotation locus of isocenter of radius 0.38 mm and an average MLC absolute positional error of 0.29 mm, consistent with use for SBRT. Specifications for clinical introduction were met.

Keywords

acceptance tests, dosimetry, magnetic field, MRI-linac, Unity, Radiation, Instrumentation, Radiology Nuclear Medicine and imaging

Citation

Woodings, S J, de Vries, J H W, Kok, J M G, Hackett, S L, van Asselen, B, Bluemink, J J, van Zijp, H M, van Soest, T L, Roberts, D A, Lagendijk, J J W, Raaymakers, B W & Wolthaus, J W H 2021, 'Acceptance procedure for the linear accelerator component of the 1.5 T MRI-linac', Journal of applied clinical medical physics, vol. 22, no. 8, pp. 45-59. https://doi.org/10.1002/acm2.13068